 +=====Pre-Examination Process=====
 +<html><span style="font-size:200%;">Sample Receiving</span></html> 
  
-__**Sample Acceptance Criteria +''Sample Acceptance Criteria''
-**__+
  
 __ __**Blood sample**__ __ __**Blood sample**__
Line 63: Line 64:
 ** **
  
-  * Unordered List ItemImproperly filled Requisition forms.+  * Improperly filled Requisition forms.
      
-  * Unordered List ItemImproperly labeled samples.+  * Improperly labeled samples.
      
-  * Unordered List ItemDetails on Requisition form is not matched with sample container details+  * Details on Requisition form is not matched with sample container details
      
-  * Unordered List ItemImproper Vacutainers +  * Improper Vacutainers 
      
-  * Unordered List ItemInsufficient sample volume / Overfilled sample+  * Insufficient sample volume / Overfilled sample
      
-  * Unordered List ItemClotted sample (Whole blood/Plasma)+  * Clotted sample (Whole blood/Plasma)
      
-  * Unordered List ItemSpill over sample.+  * Spill over sample.
      
-  * Unordered List ItemHemolyzed sample.+  * Hemolyzed sample.
      
-  * Unordered List ItemSample received after 4 hours of collection.+  * Sample received after 4 hours of collection.
      
-  * Unordered List ItemDiluted Sample+  * Diluted Sample
  
 **Urine Sample:** **Urine Sample:**
  
-  * Unordered List ItemImproperly filled Requisition forms.+  * Improperly filled Requisition forms.
  
-  * Unordered List ItemImproperly labeled samples.+  * Improperly labeled samples.
  
-  * Unordered List ItemDetails on Requisition form is not matched with sample container details+  * Details on Requisition form is not matched with sample container details
  
-  * Unordered List ItemInsufficient sample volume.+  * Insufficient sample volume.
  
-  * Unordered List ItemImproper sample container+  * Improper sample container
  
-  * Unordered List ItemSpill over sample.+  * Spill over sample.
  
 {{ :screenshot_2026-02-26_091824.png?nolink&400 |}} {{ :screenshot_2026-02-26_091824.png?nolink&400 |}}
Line 110: Line 111:
 **A.**Insufficient sample volume. **A.**Insufficient sample volume.
  
-  * Unordered List ItemDo not reject the sample.+  * Do not reject the sample.
  
-  * Unordered List ItemPerform as many examinations as possible as per priority.+  * Perform as many examinations as possible as per priority.
    
-  * Unordered List Item  Priority of testing will be decided by the laboratory incharge or consultant+  * Priority of testing will be decided by the laboratory incharge or consultant.
-Inform the customer.  +
-Enter remarks of insufficient volume in the report +
-All pediatric samples with insufficient volume will be received by the laboratory.+
  
 +  * Inform the customer.
 + 
 +  * Enter remarks of insufficient volume in the report.
  
 +  * All pediatric samples with insufficient volume will be received by the laboratory.
  
  
 +**B** Hemolyzed sample.
  
 +  * Do not analyze SGOT, SGPT, LDH, Potassium, Bilirubin, Ammonia, Hemoglobin, Hematocrit from hemolyzed samples.
  
 +  * From a less hemolyzed sample Glucose, Cholesterol, Uric acid, Urea, Creatinine, Albumin, Alkaline phosphatase, Total protein, UIBC, Sodium, Calcium can be performed.
 +
 +  * Take the decision of which parameters can be analyzed from this sample.
 +
 +  * Enter remarks of Hemolyzed samples in the report.
 + 
 +**C** If blood is collected in EDTA Vacutainer Bilirubin, Cholesterol, Triglyceride, ALT, AST, Creatinine, Urea can be performed. However it is unsuitable for Coagulation study,Enzyme assay, Hormonal assay & electrolyte analysis or methods that require electrolyte or metallic cofactors.
  
 +→ Entry of received sample and requested parameters are to be entered on software.
  
  
 +{{ :screenshot_2026-02-27_220421.png?nolink&400 |}}
  