• Switch on the main power supply.
• Switch on the STAGO switch
• Switch on the monitor
• Global verification is done automatically and on completion GLOBAL VERIFICATION DONE is displayed on monitor
• To continue press ‘ENTER’
• Text box is displayed on screen: START OF TEST; Press ‘Esc’ to continue
• “Test Panel” is displayed on monitor; Press ‘Esc’ for MENU
MENU contains: status/ loading/ files/ calibration- control/ set up/ maintenance/
• Select option ‘loading’ and move on products to load the reagents.
Product reagent drawer opens:
Reagents for PT ( STA DESORB U; STA Neoplastine CI plus 5) are first read by barcode and then product is kept at appropriate place according to size of vial
details are entered ( reagent ID, reagent name, volume, stability, lot number and position)
• Press ‘Esc’ to return to main MENU
• Select option ‘calibration- control’; select the test option from the list
• Press ‘Esc’ for the option and select “change range” and enter the reference range and save (F10)
• Press ‘Esc’ to return to the option ; enter the access code and select option “RUN”
• Press ‘Esc’ to return to the option and select “result list”
• Press ‘Esc’ to return to main MENU and select the option “Loading” ; select samples (F1); sample drawer will open. Samples are loaded by giving ID and sample position; Test is selected and validated by pressing F10
• Samples are tested automatically. When the results are displayed on screen printout taken
• Switch on the main power supply.
• Switch on the STAGO switch
• Switch on the monitor
• Global verification is done automatically and on completion GLOBAL VERIFICATION DONE is displayed on monitor
• To continue press ‘ENTER’
• Text box is displayed on screen: START OF TEST; Press ‘Esc’ to continue
• “Test Panel” is displayed on monitor; Press ‘Esc’ for MENU
MENU contains: status/ loading/ files/ calibration- control/ set up/ maintenance/ Halt
• Select option ‘loading’ and move on products to load the reagents.
Product reagent drawer opens: reagents for a PTT ( STA DESORB U; C.K.Prest and CaCl2) are first read by barcode and then product is kept at appropriate place according to size of vial
details are entered ( reagent ID, reagent name, volume, stability, lot number and position)
• Press ‘Esc’ to return to main MENU
• Select option ‘calibration- control’; select the test option from the list
• Press ‘Esc’ for the option and select “change range” and enter the reference range and save (F10)
• Press ‘Esc’ to return to the option ; enter the access code and select option “RUN”
• Press ‘Esc’ to return to the option and select “result list”
• Press ‘Esc’ to return to main MENU and select the option “Loading” ; select samples (F1); sample drawer will open. Samples are loaded by giving ID and sample position; Test is selected and validated by pressing F10
Samples are tested automatically. When the results are displayed on screen printout taken
Maintenance:
1. Check ethylene glycol level- every day
Ethylene glycol is filled upto 80 mark.
2. Needle purge – weekly
Go in maintenance then press needle purge.
Press F1, F2, F3 one by one at 2 minutes interval.