A] INTERNAL QUALITY CONTROL
1) Daily two urinalysis control samples (Positive & Negative) run on automated urine analyzer
PROCEDURE
LIMITAIONS Substances that cause abnormal urine color may affect the readability of test pads on urinanalysis reagent strips. These substances include visible levels of blood or bilirubin and drugs containing dyes, nitrofurantoin or riboflavin.
2)Daily one random patient sample is analyzed by Two/Three parameter strips and multistix strips (8 parameters/10 parameters) and by automated analyzer. Every 6 months one random urine sample is also checked by manual method. After analysis , the results are recorded in record register. CP: C/records/file/9/internal Quality control records- B.
3) Yearly twice precision check on automated analyzer. Reports of comparative observation of all strips are preserved in CP:C/records/file/9/Internal Quality control records-D
4) Weekly, Physical, chemical & microscopic examination of one random urine sample is done by two different Technicians & two different pathologists and results are compared to each other.
(Inter analyst comparison: Physical & Microscopic examination)