1. Starting Power
2. Starting software
3. Reagents placement
4. Program first run of morning
6. Placement of QC(0.5 ml)
7. Placement of Samples(0.4 ml)
8. Starting run of morning
9. Using Erba XL-640 after first run of morning:
10. Fluid levels check:
| Container | Action | Indication |
|---|---|---|
| DI Water Container | Fill upto upper mark | When level reach lower mark |
| Waste | Empty upto lower mark | When level reach upper mark |
| Cleaning solution | Replace cleaning solution Container | When level reach lower mark |
| Biohazardous waste | Empty upto lower mark and add 2 liter 4-6% HOCl | When level reach upper mark |
Make cleaning solution by adding 100 ml of Erba XL-Wash/Extran in 10 liter water
11. When
| Step | Container | Action | |
|---|---|---|---|
| 1 | DI Water Container | DIW fill upto mark | If Alarm do not stop > go to step-2 |
| 2 | Waste | Empty upto mark | If Alarm do not stop > go to step-3 |
| 3 | Cleaning solution | DIW fill upto mark + 100ml stock cleaning solution | If Alarm do not stop > go to step-4 |
| 4 | Biohazardous waste | Empty upto mark,then add [2 L HOCl] | If Alarm do not stop > call resident |