Equipment(Rapidpoint-500_siemensSerial no-2741281029) Documents and Records
RequirementType and LocationHelp
a.identity of the equipmentMettler-Toledo weight machine HOSP-INSTRU-REG. PG NO-21 NO-21 Use Equipment Name and Hospital Register page Number to create ID. It is same as in LIS cl_general.equipment.equipment
upload pdf of orders
b.manufacturer’s name, model and serial number or other unique identificationMettler-Toledo electronic weight balance Model no-PL6001E/A04 SERIAL NO-2741281029 AND SERIAL NO-2741281030 pic-1 sr_no-2741281029-1.jpg pic-2 sr_no_2741281029.jpg pic-3 sr_no_2741281030.jpg pic-4 sr_no-2741281030-2.jpgUnique ID of equipment in manufacturer's context
c.contact information for the supplier or the manufacturerDocument in Dokuwiki upload records in lis under equipment menuAddress, company/local sale/engineer phone/email, number for automated complaint registration etc
d.date of receiving and date of entering into service02-09-2020Same as in register
e.locationResearch labLab section/room etc
f.condition when received (e.g. new, used or reconditioned)New
g.manufacturer’s instructionsmeeler_toleido_user_manual.pdfUser manual pdf
h.Procedural document for and records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratory it was received in biochemistry lab on 2-9-2020 and after receiving we opened both instrument and seen all parts of the machine During that it was noticed that 1 part(cap) of both instrument was absent and that was informed to supplier calibration was done for acceptability For serialr no 2741281030 with 2.0 kg=2000.1 gm For 1.0 kg=1000.0 gm For 10gm=10.0 gm and For 1.0 gm=1.0gm ,For serial no 2741261029 with 2.0 kg=2000.2 gm For 1.0 kg=1000.1 gm For 10.0 gm=10.0gm For 1.0 gm=1.0 gm.It was noticed after reading it user manual that while machine wis not in use pput cap on measuring part and keep aside its upeer plate as it will measure that plate weight and machine will disfunction in short time so we had put that upper plate separate to prevent early disfunctionmetller_toledo_bill.pdfIt may be in user manual. In that case write chapter/page numbers for easy reference
i.Record of maintenance carried out and the Procedural(schedule) document for preventive maintenance
j.Procedural document and records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verifications
k.Documented Procedure for repair and Records of damage to, or malfunction, modification, or repair of the equipment