Conclusion

  • Glucose reagent interferes with TBil estimation
  • Probably there is reagent carry over.
  • Introduce one Deteregent wash cycle in between Glucose (interfering test) and Tbil (Interfered Test) in ERBA 640 XL.
  • Documentation for introducing Wash Cycle

Mechanism

  • H2O2 produced by GOD POD reaction destroys bilirubin.
  • So Bilirubin was low in result. One wash cycle was set inbetween two tests to eliminate carry over.