User Tools

Site Tools


Sidebar

Dr Piyush B. Tailor

Be Human - Be Part of Change

Lectures

Model Questions

Question Papers

Moodle
Social Media

About

nchsls:c:biochemistry:document:intereference_by_god_pod_glucose_reagent

Conclusion

  • Glucose reagent interferes with TBil estimation
  • Probably there is reagent carry over.
  • Introduce one Deteregent wash cycle in between Glucose (interfering test) and Tbil (Interfered Test) in ERBA 640 XL.
  • Documentation for introducing Wash Cycle

Mechanism

  • H2O2 produced by GOD POD reaction destroys bilirubin.
  • So Bilirubin was low in result. One wash cycle was set inbetween two tests to eliminate carry over.
nchsls/c/biochemistry/document/intereference_by_god_pod_glucose_reagent.txt · Last modified: 2022/08/13 07:34 (external edit)