Take 5 ml Qc water(DI water of 0.00 microsiemens/cm) in volumetric flask.
Gently transfer water in Randox Human assay level 2 and level 3 bottle.Pour last drop of water by simply rotate flask over edge of bottle.Do not beat the flask over bottle.
Mix it by gentle inversion and eversion.
Put it at room temperature for 30 minute and watch for all powder particle has been dissolved.
Fill 10 ependroff cup of 500 microliter.and seal with it parafilm.
Preparation of quality control samples for lipase
Make low and high lipase serum pool as quality control samples.
For low lipase serum pool , collect leftover serum from OPD samples.
Find serum lipase > 20 IU/L for 3-4 days.
Collect sample in 10 ml vacutte everyday.kept vacutte in deep freeze.
After approximately 10 ml serum collected,Centrifuge vacutte at 3000 RPM for 10 minutes.
Fill 100 microliter in ependroff cup. Make such 90 such ependroff cups.
Preparation of quality control samples for Microprotein
wash 50 ml flask.
take 500 microliter of calibrator1) in that flask.
take DI water in flask upto the mark and wrap parafilm over mouth of flask.
mix it well.
wash 50 ml bottle and fill this diluted calibrators in flask in washed bottle.
make approximately 100 cup of MPR calibratorwith 500 microliter volume and airtight cup with parafilm.
put all cups in deep fridge (-40oc).
Note: while making cup of 500 microliter shack 50 ml bottle with diluted calibrators intermittantly.
Preparation of quality control samples for ADA
Make low and high ADA pool as quality control samples.
Find ADA < 15 and ADA > 15 for 3-4 days.
Collect sample in 10 ml vacutte everyday.kept vacutte in deep freeze.
After approximately 10 ml serum collected,Centrifuge vacutte at 3000 RPM for 10 minutes.
Fill 100 microliter in ependroff cup. Make such 90 such ependroff cups.