skip to content
Dr Piyush Tailor
User Tools
Log In
Site Tools
Search
Tools
Edit this page
Old revisions
Backlinks
Recent Changes
Media Manager
Sitemap
Log In
>
Recent Changes
Media Manager
Sitemap
Trace:
•
5_lugol_s_iodine
•
loading_reagents_in_erba_xl-640
•
sample_retention_in_xl-1000
•
how_to_send_qaf_results
•
calibrator_and_qc_targets
•
method_of_washing_cups_for_qc_and_calibrators
•
iqc_of_vitros_3600
•
how_to_prepare_heparinized_syringe
•
total_bilirubin_reagent
•
erba_xl-640_fluid_managing
Sidebar
Dr Piyush B. Tailor
Be Human - Be Part of Change
*Home
Dr Piyush's Home
Biochemistry Home
GMC Surat Home
*Lectures
Biochemistry
Medical Laboratory Technician
Medicine
Model Questions
Biochemistry
Medicine
Optometry
Question Papers
MBBS Biochemistry
BPT Biochemistry
MLT All Subject
Social Media
Telegram
Youtube
About
Publication
Experience
Gallery
nchsls:c:biochemistry:document:erba_xl-640_fluid_managing
For Refilling DI Water and Cleaning solution
Remove sensor cap of DI water container behind Erba XL-640.
Empty DI water carba present on platform into DI water container.
close sensor cap of DI water container.(For Cleaning solution add 100 ml CLEANING SOLUTION in 10 L container and fill rest of carba with DI water.)
Edit
For Emptying Waste (Non hazardious and bio hazardious)
Remove sensor cap of Waste water container .
put Green pipe of Motor in container .
start motor power supply and empty waste.
Take out green pipe from waste container and put in tap water container for 1 min.
Switch Off the motor power supply
close sensor cap of waste container.(for Bio hazardious, add 2 L Hypochlorite in container).