the first thing that incoming tap water encounters, and as the name implies it filter out sediment.
5 um filter is used in our laboratory.
Carbon filter types:
RO membrane is a semipermeable membrane, allowing water to pass through while rejecting contaminants that are too large to pass through the tiny pores inthe membrane.Reverse osmosis membrane is of microscopic size (0.0001 micron)allowing only pressurized water molecules to pass and bigger molecules are rejected . Size of smallest bacterium is 0.02 um
RO membrane removes 95-99% of organics , bacteria and other particulate matter and 90-97% of all ionized and dissolved minerals but fewer of gaseous impurities
(RSO3)H + Na+ —–> (RSO3)Na + H+
(RNR3)OH + Cl- ——> (RNR3)Cl + OH-
Mixed bed deionizers usually produce water with a specific resistance that exceeds 10 Mohm/cm.
Polyethersulfone (PES) membranes are hydrophilic filters constructed from pure polyethersulfone and available in pore sizes raing from 0.03 micron to 5.0 micron.
UV Filter is incorporated so as to kill any live bacteria, spores and viruses
Used to measure conductivity of the water from the output tube measured in microsiemens/cm or as TDS in ppm
CONDUCTIVITY IS RECIPROCAL OF RESISTIVITY
| TYPE 1 | TYPE 2 | TYPE 3 | |
| Microbiological content, colony forming unit/mL, cfu/mL (maximum) | 10 | 103 | NA |
| pH | NA | NA | 5.0-8.0 |
| Resistivity Mohm- cm, 25'C | 10 (in line) | 2.0 | 0.1 |
| Silicate, mg SiO2 / L(maximum) | 0.05 | 0.1 | 1.0 |
| Particulate matter | Water passed through 0.2 um filter | NA | NA |
| Organics | Water passed through activated carbon | NA | NA |
Microbiological content : The microbiological content of viable organisms , as determined by total colony count after incubation at 36 ± 1 'C for 14 hrs followed by 48 hrs at 25 ± 1 'C and reported as colony forming unit/mL (cfu/mL)
Specific resistance or resistivity: The electrical resistance in ohms measured between opposite faces of a 1.00 cm cube of an aqueous solution at a specified temperature . For these specifications , the resistivity will be corrected for 25'C and reported in Mohm/cm. The higher the amount of ionizable materials, the lower the resistivity and higher the conductivity.
Particulate matter : When water is passed through a membrane filter with a mean pore size of 0.2 um, it is considered to be free of particulate matter.
Organic material : When water is passed through a bed of activated carbon, it is considered to contain minimum organic material .
QUALITY , USE AND STORAGE OF REAGENT GRADE WATER
Type III :
Type II :
Type I :