kit literature

D-DIMER Calibration with linearity establishment

ddimer_plan.ods

  • lyophilized D-DIMER calibrator bottle is reconstituted with 1 ml(1 gm qc water) DI water, incubated for 10 to 15 minutes with periodic shaking until solutes becomes completely dissolved.
  • 200 Ul of the mixture(reconstituted calibrator) is added with 200 Ul of Di water.
  • from there again 200 Ul of solution is added to the next alliquote and 200 Ul water is added for 3 more subsequent times. (serial dilution)
  • For establishing linearity a total of 6 samples are required.
  • 1 sample is the plain calibrator(stock d-dimer), and the other one is DI water.
  • the samples are analysed in the analyser and od is plotted in the graph.
  • from there the factors are extracted and feeded in the machine.
  • For making an (IN use)ready to use reagent, 2000 uL of R1 is added with 500 uL of R2, which was already kept in fridge 2* to 8* C.
  • special wash reagent (0.1 n NAOH) is also added for wash

Calibrator traceability certificateqdd1026.pdf

sample comparison with Coral kit and Labcare kit d-dimer_sample_comparison05_08.ods

How to make D-DIMER Internal Quality Control