Frequency of calibration and IQC

Erba XL-1000
ParameterCalibrationIQCCalibration PolicyCorrelation FactorEQC Policy
calibrator and Replicates/QC and Replicates
prepare 200ul calibratuo cups and devide 500ul QC cups in to 200ul
perform recalibration of same lot if qc results are out of 1SD on same side for more than 15days
GLCWHEN LOT CHANGEs or if calibration is older than 1 month even if lot is same2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
Sodiumwhenever electrodes changed.Do 2 level IQC 5 times when fluid pack refilled and 2 level IQc at 8 AM ,3 PM , 10 PM5 calibrator CUPS X 2 time replicate and 5 IQC cups x 2 time replicatey=axRIQAS_monthly
Potassiumwhenever electrodes changed.Do 2 level Qc 5 times when fluid pack refilled and 2 level Qc at 8 AM ,3 PM , 10 PM5 calibrator CUPS X 2 time replicate and 5 QC cups x 2 time replicatey=axRIQAS_monthly
TBILWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM , 10 PMUse freshly prepared calibrators and quality control(after one hour of reconstitution)3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
DBILWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM , 10 PMUse freshly prepared calibrators andUse freshly prepared calibrators and3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicateUse TBIL Factor achieved in calibrationRIQAS_monthly
ALTWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
TGWHEN LOT CHANGES or if calibration is older than 1 month even if lot is same2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
CHOWHEN LOT CHANGES or if calibration is older than 1 month even if lot is same2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
UREAWHEN LOT CHANGES 2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
UAWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
CKMBFixed Factor,adjust factor according to dilution factor when kit/protocol changed based on molar absorptivity of NADPHKIT CONTROL WHEN BOTTLE CHANGE.2 LEVEL IQC AT 8 AM, 1 LEVEL IQC AT 3 PM, 1 LEVEL IQC AT 10 PM NA.Fixed Factor
ALBWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatedy=ax+bRIQAS_monthly
ALPWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM ,10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
TPWHEN LOT CHANGES2 LEVEL IQC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 IQC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
AMYWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
CALWhen Lot changes because reagent is used up within 35 days2 LEVEL QC at 8 AM ,3 PM , 10 PM (In duplicate and second value is considered)3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
CAL(OCPC)Every four hourly2 LEVEL QC 4 Hourly and with each new batch(In duplicate and second value is considered)4 calibrators(HAlf cal 2, half cal3, cal 2, cal 3) cups X 2 time replicate, one blank x 2 times y=ax+bRIQAS_monthly
CRWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM ,10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 albumin(5%) blank cups X 4 time replicatey=ax+b(Dont Use the value of blank for finding the equation)RIQAS_monthly
LDHWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
LIPFixed Factor(-500)2 level pooled sera at 8 amNALiterature factor=500
IRONCalibration when lot changes2 Level QC with each batch3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
UIBCCalibration when lot change and along with each batch. each batch calibration is minicalibration.1 Level QC with each batch3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax+bRIQAS_monthly
ADAin every batch, with 60 U/ml ADA standard X 3 times replicate Two level pooled sera in every batchNA QA Forum
MPRWhen Lot changestwo level diluted(1:100 times) calibrator sera at 8 am (In duplicate and second value is considered)10 Calibrators cups X 2 time replicate and 5 blank cups X 2 time replicatey=axCSF_RIQAS_monthly
CHOHFixed factor(caculated from CHO factor),change it when factor of cholesterol changes2 LEVEL QC dailyNA(cholesterol factor)x1.1RIQAS_monthly
CHE whenever new 30ml bottle is reconstituted 2 LEVEL QC AT 8 AM, 1 LEVEL QC AT 3 PM, 1 LEVEL QC AT 10 PM 4 calibrators cups X 2 time replicate and 4 QC cups X 2 time replicate and 2 blank cups X 2 time replicate y=axRIQAS_monthly
PHOWhen Lot changes2 LEVEL QC at 8 AM ,3 PM , 10 PM(In duplicate and second value is considered)3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
Dithionite TestQualitative test Two level control (positive and negative) run in each batchNA QA Forum_monthly
Hemoglobin electrophoresis-Detection of HbA, HbS & HbFQualitative test Control of HbA, HbS and HbF are run with each batchNA QA Forum_monthly
Fluid Glucosesame as serum glucose Same as serum Glucose IQC3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA FORUM 6 monthly and in CSF RIQAS monthly
Fluid Protein (Biuret)Same as serum proteinSame as Serum protein IQC 3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_6 monthly
Fluid Protein (Pyrogallol Red)Same as MPR Two level pooled fluid in every batch10 serum Calibrators cups X 2 time replicate and 5 blank cups X 2 time replicatey=axQA FORUM_6 monthly
Serum Ketone bodies Qualitative TestTwo level pooled sera( Normal and Acetone spiked) Twice a dayNA QA Forum_6 monthly
Urine Ketone bodies Qualitative Test Same as serum ketone NA QA Forum_6 monthly
Urine CalciumSame as serum calcium calibration 1 level pooled spiked Acidify urine once when test done 3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_6 monthly
Urine CreatinineSame as serum creatinine 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_6 monthly
Urine PhosphorousSame as serum Phosphorus 1 level pooled spiked Acidify urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_6 monthly
Urine SodiumWhen Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_6 monthly
Urine Protein (Biuret)When Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_6 monthly
Urine Protein (Pyrogallol Red)When Lot changes 2 level diluted calibrator once a day10 serum Calibrators cups X 2 time replicate and 5 blank cups X 2 time replicatey=axQA Forum_6 monthly
Urine Uric Acid (Uricase POD)same as serum Uric Acid 1 level pooled spiked basic urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axQA Forum_monthly
Protein electrophoresis-Detection of M bandQualitative Test Three level QC run with each batchNA QA Forum_6 monthly
CK(IFCC at 37)When lot changes2 LEVEL QC AT 8 AM, 1 LEVEL QC AT 3 PM, 1 LEVEL QC AT 10 PM1 kit standard cup and 1 blank cup 5times replicatedy=ax+bRIQAS monthly
CRP(turbidimetry)When lot changed2 level in-house pool in morning 8 amserial dilution of kit standard how_to_do_serial_dilution_of_crp_standard.ods 5 cups with 4 replicatesy=ax+bQA Forum 6 monthly
DDIMER(turbidimetry)When lot changed1 level diluted kit standard in morning 8 am serial dilution of kit standard 5 cups with 1replicatehow_to_do_serial_dilution_of_ddimer_standard.odsy=axQA Forum 6 monthly
Total 25-OH Vit-D(ECLIA)When lot changedIQC weekly one levelTwo level calibrator freeze dried form reconstitute with 1ml qc waterAutocalibratedRIQAS monthly
Ferritin When lot changed or Calibration expiredIQC weekly one levelThree level calibrator freeze dried form reconstitute with 1ml qc waterAutocalibratedRIQAS monthly
Procalcitonin When lot changed or Calibration expiredIQC weekly one levelTwo level calibrator ready to use(Frozen sample)AutocalibratedQA Forum 6 monthly
TSH When lot changed or Calibration expired2 level IQC daily at 8AM, 1 level at 3 PM, 1 level at 10 PMThree level calibrator ready to useAutocalibratedRIQAS monthly
FT3 When lot changed or Calibration expiredIQC weekly one levelThree level calibrator freeze dried form reconstitute with 1ml qc waterAutocalibratedRIQAS monthly
FT4 When lot changed or Calibration expiredIQC weekly one levelThree level calibrator freeze dried form reconstitute with 1ml qc waterAutocalibratedRIQAS monthly
LH When lot changed or Calibration expiredIQC weekly one levelTwo level calibrator ready to use(Frozen sample)AutocalibratedRIQAS monthly
FSH When lot changed or Calibration expiredIQC weekly one levelTwo level calibrator ready to use(Frozen sample)AutocalibratedRIQAS monthly
Prolactin When lot changed or Calibration expiredIQC weekly one levelTwo level calibrator freeze dried form reconstitute with 1ml qc waterAutocalibratedRIQAS monthly
NTProBNP2 When lot changed or Calibration expiredIQC weekly one levelThree level calibrator ready to useAutocalibratedNot done
HsTnI When lot changed or Calibration expiredIQC weekly one levelThree level calibrator ready to useAutocalibratedQA Forum 6 monthly
B-HCG When lot changed or Calibration expiredIQC weekly one levelThree level calibrator freeze dried form reconstitute with 1ml qc waterAutocalibratedRIQAS monthly
Vit B12 When lot changed or Calibration expiredNot done because no sample received Two level calibrator ready to use(Frozen sample)AutocalibratedRIQAS monthly
CoV2G When lot changed or Calibration expiredNot doneSingle level calibrator ready to useAutocalibratedNot done
Lactate One point autocalibration half hourly, two point autocalibration at 2 hourly one level IQC dailyOne point autocalibration half hourly, two point autocalibration at 2 hourly-Not done
HCO3 One point autocalibration half hourly, two point autocalibration at 2 hourly one level IQC dailyOne point autocalibration half hourly, two point autocalibration at 2 hourly-Not done
pH One point autocalibration half hourly, two point autocalibration at 2 hourly one level IQC dailyOne point autocalibration half hourly, two point autocalibration at 2 hourly-Not done
pCO2 One point autocalibration half hourly, two point autocalibration at 2 hourly one level IQC dailyOne point autocalibration half hourly, two point autocalibration at 2 hourly-Not done
pO2 One point autocalibration half hourly, two point autocalibration at 2 hourly one level IQC dailyOne point autocalibration half hourly, two point autocalibration at 2 hourly-Not done
HbA1CWeekly and When buffer changed2 level IQC from in house pool with each batch.Two level calibrator freeze dried form reconstitute with 4ml qc water-QA Forum 6 monthly
Erba Chem-5 Plus V2
ParameterCalibration IQC
When in use(Erba XL-640 breakdown)When not in useWhen in use(Erba XL-640 breakdown)When not in use
GLCEvery batchEvery month before External QC and when lot changes2 level QC With Every batch2 level QC every Sundayy=ax
BILwhen lot changesEvery month before External QC and when lot changes2 level QC 3 times a day2 level QC every sundayy=ax
CRwhen lot changesEvery month before External QC and when lot changes2 level QC 3 times a day2 level QC every sundayy=ax
ALTwhen lot changesEvery month before External QC and when lot changes2 level QC 3 times a day2 level QC every sundayy=ax
ADAin every batchNot applicableTwo level pooled sera in every batchNot applicable