—CRP kit literature pdf— crp_kit_lit.pdf —CRP Calibration with linearity establishment— crp_plan.ods

  • lyophilized crp bottle is reconstituted with 1 ml DI water, incubated for 10 to 15 minutes with periodic shaking until solutes becomes completely dissolved.
  • 100 Ul of the mixture is added with 100 Ul of Di water.
  • from there again 100 Ul of solution is added to the next alliquote and 100 Ul water is added for 3 more subsequent times. (serial dilution)
  • For establishing linearity a total of 6 samples are required.
  • 1 sample is the plain calibrator(as CRP stock), and the other one is DI water(as blank).
  • the samples are analysed in the analyser and od is plotted in the graph.
  • from there the factors are extracted and feeded in the machine.
  • the R1 volume is change from 200 UL to 160 ul and r2 volume is changed from 50 to 40 ul.
  • special wash reagent (0.1 n NAOH) is also added for wash