*Learn operation of Erba XL-640 and Erbachem.
| references | Cholinesterase Kit literature of RANDOX |
| Note: When a new lot is opened, verify kit literature with the uploaded version. | |
| Upload newer version when required and make necessary changes in procedure and process |
| environmental and safety controls | MSDS of CHE Randox Kits |
| The examination is done on Erbachem and Erba XL-640 |

Note: Following instruction is useful only if operator is familiar with Erbachem
| Parameter | Value |
|---|---|
| Sample volume(ul) | 10 |
| Reagent volume(ul) | 775 |
| Filter | 405 nm |
| Aspitation Volume(ul) | 500 |
| Type of reaction | 2 point |
| Decimal place | 1 |
| Delay time | 30sec |
| Reading time | 60sec |
| Calibration, Result calculation, Quality control and sample examination in Erbachem | |
|---|---|
| Prepare for calibration as below when lot changes or there is relevent NC. Refer Calibrator and QC | |
| testing material | Replicates |
| Water blank | 2 |
| Normal Range Calibrator | 4 |
| Abnormal range Calibrator | 4 |
| Normal Range Control | 4 |
| Abnormal range Control | 4 |