General Instruction

Centrifuge Remi PR-23 HOSP-EQ-P-138-VCDB-4173

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-138-VCDB-4173
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi PR-23 VCDB-4173
contact information for the supplier or the manufacturerRANJIT-9824955357,CHETAN-9998887382,PRGNESH-9974564999
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service09-06-2014
locationClinical Biochemistry Lab
condition when received (e.g. new, used or reconditioned)New
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Centrifuge Remi PR-23 HOSP-EQ-P-138-VCDB-4174

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-138-VCDB-4174
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi PR-23 VCDB-4174
contact information for the supplier or the manufacturerRANJIT-9824955357,CHETAN-9998887382,PRGNESH-9974564999
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service17-06-2011
locationClinical Biochemistry Lab
condition when received (e.g. new, used or reconditioned)New
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Centrifuge Remi PR-23 HOSP-EQ-P-138-VCDB-4175

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-138-VCDB-4175
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi PR-23 VCDB-4175
contact information for the supplier or the manufacturerRANJIT-9824955357,CHETAN-9998887382,PRGNESH-9974564999
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service17-06-2011
location Biochemistry Research Lab
condition when received (e.g. new, used or reconditioned)New
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Centrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25664

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25664
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi-PR-23 ZEHN 25664
contact information for the supplier or the manufacturerMr.Rajesh Tiwari- 9136945168
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service17-08-2018
location Biochemistry Research Lab
condition when received (e.g. new, used or reconditioned)New
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Centrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25665

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25665
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi-PR-23 ZEHN 25665
contact information for the supplier or the manufacturerMr.Rajesh Tiwari- 9136945168
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service17-08-2018
location Clinical Biochemistry Laboratory
condition when received (e.g. new, used or reconditioned)RC copy51-e-148-02052018.pdf
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Centrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25666

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25666
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi-PR-23 ZEHN 25666
contact information for the supplier or the manufacturerMr.Rajesh Tiwari- 9136945168
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service17-08-2018
location Biochemistry Research Lab
condition when received (e.g. new, used or reconditioned)New
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Centrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25667

Equipment Documents and Records
identity of the equipmentCentrifuge Remi PR-23 HOSP-EQ-P-52-ZEHN 25667
manufacturer’s name, model and serial number or other unique identificationCentrifuge Remi-PR-23 ZEHN 25667
contact information for the supplier or the manufacturerMr.Rajesh Tiwari- 9136945168
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service17-08-2018
location Clinical Biochemistry Laboratory
condition when received (e.g. new, used or reconditioned)RC copy51-e-148-02052018.pdf
manufacturer’s instructionsuser_manual_centrifuge_pr-23
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php

Research Compufuge Remi PR-24 HOSP-EQ-P-51-ZCBN-04721

Equipment Documents and Records
identity of the equipmentResearch Compufuge Remi PR-24 HOSP-EQ-P-51-ZCBN-04721
manufacturer’s name, model and serial number or other unique identificationResearch Compufuge Remi PR-24 ZCBN-04721
contact information for the supplier or the manufacturer
contact information for the maintanenceRajesh Tiwari-7028675731,Email:service@remilabworld.com
date of receiving and date of entering into service02-06-2017
location Clinical Biochemistry Laboratory
condition when received (e.g. new, used or reconditioned) installation_report_and_purchase_order.pdf3.at_of_pr-24.pdf
manufacturer’s instructionsnew_centrifuge_user_manual.pdf
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryhttp://12.207.3.240/equipment_record.php
maintenance carried out and the schedule for preventive maintenancehttp://12.207.3.240/equipment_record.php
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationshttp://12.207.3.240/equipment_record.php
damage to, or malfunction, modification, or repair of the equipmenthttp://12.207.3.240/equipment_record.php
Do in house of calibration of centrifuge every 6 monthlyProcedure of calibration calibration_of_centrifuges.odt

How To Operate Remi Centrifuge PR 23

It was seen that During centrifugation vacuttee cap touches rotor of centrifuge while it is in horizontal position so it prevent the bucket to come in vertical position.to prevent this following should be done.during cotton filling equal amount of cotton should be fill in both the rotor(opposite one)in same position.see below picture:

STEPS:

  1. Load balanced samples and close the lid
  2. After power on check following on Display[PRG. No=0] [RPM=3000] [Rotor No=19] [Brake=M] [Min=10]
  3. Press START button
  4. Wait till alarm (dip noice) for 5 times and again RPM = 3000
  5. Open the lid and unload samples

Precautions:

  1. Do not change the parameters of programs
  2. Press the button of lid lock of cover at the time of closing cover to prevent malfunction of lid lock
  3. Match bucket number with number on rotor,both numbers facing each other.
  4. Do not handle without gloves.
  5. Stay away during centrifugation.
  6. Switch off the centrifuge when not in use.

centrifugation

Steps

  1. Load balanced samples and close the lid
  2. After power on check following on Display
  3. [PRG. No=0] [RPM=3000] [Rotor No=19] [Brake=M] [Min=10]
  4. Press START button
  5. Centrifuge fluoride and plain vacutte for 10 minutes at 3000 rpm.
  6. Wait till alarm (dip noice) for 5 times and again RPM = 3000
  7. pen the lid and unload samples

Mistakes

  1. Do not change the parameters of programs
  2. Press the button of lid lock of cover at the time of closing cover to prevent malfunction of lid lock
  3. Match bucket number with number on rotor,both numbers facing each other.
  4. Do not handle without gloves.
  5. Stay away during centrifugation.

Observations:

  1. If sample is highly hemolyzed,lipemic or icteric ,add examination interference in that sample id and write “sample is hemolyzed”.
  2. If sample is highly hemolyzed , add examination rejection in that sample id and write “Sample is highly hemolyzed send new sample.
  3. If any fluid is highly viscous so that it can block the probe, add examination rejection in that sample id and write “sample is too viscous for any analysis to be done”

How to clean centrifuge: Centrifuge Maintenance sheetcentrifuge_maintenence_sheet.ods

  1. First remove all 4 buckets.Pull bucket vertically to remove it.
  2. ,
  3. Clean buckets and entire centrifuge(internal part) with 1% HOCL solution with cotton.
  4. Clean buckets and entire centrifuge(internal part) with cotton soaked with tap water.
  5. In case of spillage with blood,
  6. keep cotton soaked with 1% HOCL over spillage for 30 minutes.
  7. Discard that cotton according to waste management rule.
  8. Again Clean buckets and entire centrifuge(internal part) with 1% HOCL solution with cotton.
  9. Clean buckets and entire centrifuge(internal part) with cotton soaked with tap water.

Precautions:

  1. Be careful about exchange of samples.
  2. Do not use same tip to separate the serum. Use new tip for every sample.

What to do when our centrifuge get troubled while run??centrifuge_get_trouble.odt}}

-There is error in centrifuge,stop after reach 3000 RPM ,then check by removing rotor still error continue

-Then again properly feet rotor, error solved centrifuge start working properly

Deep freeze maintenance sheet