nchsls:c:biochemistry:document:calibration_and_quality_control_frequency

Frequency of calibration and IQC

Erba XL-640
ParameterCalibrationIQCCalibration PolicyCorrelation FactorEQC Policy
calibrator and Replicates/QC and Replicates
prepare 200ul calibratuo cups and devide 500ul QC cups in to 200ul
perform recalibration of same lot if qc results are out of 1SD on same side for more than 15days
GLCWHEN LOT CHANGEs or if calibration is older than 1 month even if lot is same2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
Sodium, PotassiumTwo level when batch is started, one level with each sample(Inbuilt)Do 2 level Qc 5 times when fluid pack refilled and 2 level Qc at 8 AM ,3 PM , 10 PM3 calibrator CUPS of 1 ml X 5 times replicate and 3 QC cups of 1ml x 5 times replicate. After running the cups do check that around 500 ul sample is left in the cups which can be reused later. the Due to difference in expected dimension of HITACHI cup and real dimension of eppendorf cup, aspiration of 70ul is error prone if residual volume is less than 400 ulRIQAS_monthlyWhy in measurement of electrolyte do we take more than 400 ul serum for doing replication ? hitachi_cups.odt
TBILWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PMUse freshly prepared calibrators and quality control(after one hour of reconstitution)3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
DBILWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PMUse freshly prepared calibrators andUse freshly prepared calibrators and3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
ALTWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
TGWHEN LOT CHANGES or if calibration is older than 1 month even if lot is same2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
CHOWHEN LOT CHANGES or if calibration is older than 1 month even if lot is same2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
UREAWHEN LOT CHANGES 2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
UAWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
CKMBFixed Factor,adjust factor according to dilution factor when kit/protocol changed based on molar absorptivity of NADPHKIT CONTROL WHEN BOTTLE CHANGE.when 30 ml bottle will reconstituted.Fixed Factor
ALBWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatedy=axRIQAS_monthly
ALPWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM ,10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
TPWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
AMYWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
CALWhen Lot changes because reagent is used up within 35 days2 LEVEL QC at 8 AM ,3 PM , 10 PM (In duplicate and second value is considered)3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
CRWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM ,10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 albumin(5%) blank cups X 4 time replicatey=ax+bRIQAS_monthly
LDHWHEN LOT CHANGES2 LEVEL QC at 8 AM ,3 PM , 10 PM3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
LIPFixed Factor(-500)2 level pooled sera at 8 amNALiterature factor=500
IRONCalibration when lot change and along with each batch2 Level QC with each batch3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
UIBCCalibration when lot change and along with each batch2 Level QC with each batch3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
ADAin every batch, with 60 U/ml ADA standard X 3 times replicate Two level pooled sera in every batchNAQA Forum
MPRWhen Lot changestwo level diluted(1:100 times) calibrator sera at 8 am (In duplicate and second value is considered)10 Calibrators cups X 2 time replicate and 5 blank cups X 2 time replicatey=axRIQAS_monthly
CHOHFixed factor(caculated from CHO factor),change it when factor of cholesterol changes2 LEVEL QC EVERY MORNING AND EVENINGNA(cholesterol factor)x1.1RIQAS_monthly
CHE whenever new 30ml bottle is reconstituted RIQAS_monthly
CKWhen Lot changes1 level QC every morning 3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
PHOWhen Lot changes2 LEVEL QC at 8 AM ,3 PM , 10 PM(In duplicate and second value is considered)3 calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=axRIQAS_monthly
Dithionite TestQualitative test Two level control (positive and negative) run in each batchNA RIQAS_monthly
Hemoglobin electrophoresis-Detection of HbA, HbS & HbF Control of HbS and HbF are run with each batchNA
Fluid GlucoseWhen Lot changes Two level pooled fluid in every batch3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Fluid Protein (Biuret)When Lot changes Two level pooled fluid in every batch3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Fluid Protein (Pyrogallol Red)When Lot changes Two level pooled fluid in every batch10 serum Calibrators cups X 2 time replicate and 5 blank cups X 2 time replicatey=ax
Serum Ketone bodies Qualitative TestTwo level pooled sera( Normal and Acetone spiked) once a dayNA
Urine Ketone bodies Qualitative TestTwo level pooled urine ( Normal and Acetone spiked) once a dayNA
Urine CalciumWhen Lot changes 1 level pooled spiked urine once when test done 3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Urine CreatinineWhen Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Urine PhosphorousWhen Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Urine SodiumWhen Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Urine Protein (Biuret)When Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Urine Protein (Pyrogallol Red)When Lot changes 1 level pooled spiked urine once when test done10 serum Calibrators cups X 2 time replicate and 5 blank cups X 2 time replicatey=ax
Urine Uric Acid (Pyrogallol Red)When Lot changes 1 level pooled spiked urine once when test done3 Serum calibrators cups X 4 time replicate and 3 QC cups X 4 time replicate and 3 blank cups X 4 time replicatey=ax
Protein electrophoresis-Detection of M band Two level QC run with each batchNA
Erba Chem-5 Plus V2
ParameterCalibration IQC
When in use(Erba XL-640 breakdown)When not in useWhen in use(Erba XL-640 breakdown)When not in use
GLCEvery batchEvery month before External QC and when lot changes2 level QC With Every batch2 level QC every Sundayy=ax
BILwhen lot changesEvery month before External QC and when lot changes2 level QC 3 times a day2 level QC every sundayy=ax
CRwhen lot changesEvery month before External QC and when lot changes2 level QC 3 times a day2 level QC every sundayy=ax
ALTwhen lot changesEvery month before External QC and when lot changes2 level QC 3 times a day2 level QC every sundayy=ax
CKMBFixed factorFixed factorWhen new reagent bottle reconstituted, kit control should be measuredKit control when patient sample is to be done
CHEWhen lot changesNot applicableNormal or abnormal Qc with patient sampleNot applicable
ADAin every batchNot applicableTwo level pooled sera in every batchNot applicable
nchsls/c/biochemistry/document/calibration_and_quality_control_frequency.txt · Last modified: 2022/08/13 07:34 (external edit)