nchsls:c:biochemistry:document:action_taken_when_1_3s_or_2_2s_rule_violeted

general guideline

  1. 1(3S): Random and Systemic error
  2. 2(2S): Systemic error
  3. Random error: error in precision
  4. Systemic error: error in accuracy

Look for following non-exhaustive list of ERROR-CENTERS

Integrity of QC sera
  1. Refrigerator temperature chart
  2. New QC Bottle
  3. Pipette calibration/ malfunction
  4. Interview technician who reconstituted QC sera
  5. Verify QC-Level(Normal, High), Lot number
  6. Compare with QC results of other examinations
Integrity of reagents
  1. Interview technician who mixed, refilled reagents
  2. Visually inspect reagents
  3. Compare with reagent from freshly refilled reagent-bottle
  4. Verify Manufacturer, Lot number
  5. Verify due date of calibration
  6. Compare with QC results of other examinations
Integrity of Equipment
  1. Compare with QC results of other examinations
  2. Automated chemistry analysers: Light sources, Cuvette washing, Probe Cleaning, Wash solution dispensing
  3. ISE analyses: Electrode response
ROOT CAUSE ANALYSIS FOR IQC