Haematology (IQC plan)
Hemoglobin Two level quality control morning and evening daily
Total WBC count Two level quality control morning and evening daily
Differential WBC count (Neutrophil, lymphocyte, Eosinophil, Basophil, Monocyte/others)Weekly one smear for differential count is examined by three different observers.

Daily quality control morning and evening on five part hematology analyzer
RBC count Two level quality control morning and evening daily
MCV Two level quality control morning and evening daily
MCH Two level quality control morning and evening daily
MCHC Two level quality control morning and evening daily
RDW(CV) Two level quality control morning and evening daily
HCT Two level quality control morning and evening daily
Platelet count Two level quality control morning and evening daily
PS MP Weekly one smear for MP is examined by three different observers.
Malarial antigen detection test Monthly one result of Malarial antigen detection test is compared with manual smear examination.
Peripheral smear examination Weekly one smear for PSCM is examined by three different observers.
ESR Retesting of same sample at interval of 6 hours daily on automated ESR analyzer and comparison of results between automated ESR analyzer and manual method daily.
Reticulocyte count Weekly one smear for RC is examined by three different observers.
Bone marrow smear examination One bone marrow smear is observed by two different observers every 3 monthly.
Prothrombin time Two level quality control in morning daily
Activated partial thromboplastin time. Two level quality control in morning daily
HbA2 Two level daily Quality control in morning at starting of new batch
HbF Two level daily Quality control in morning at starting of new batch