DAILY MAINTENANCE
HOW TO SEE NUMBER OF TEST REMAINING
Go to reagent menu

See the number of test remainning from
count menu

HOW TO LOAD REAGENT
Go to reagent menu

click on load/unload button

and now load the reagent
Do not forget to close the reagent tray cover

And see the status of reagent -
in use or current should be there

In case of bubble unload the reagent and again load it
HOW TO REMOVE EMPTY REAGENT
Go to reagent menu
Select the reagent you want to remove
And now click on load/unload button
Now remove reagent from reagent tray
Do not forget to close the reagent tray cover
HOW TO SEE SIGNAL REAGENT
Go to reagent menu
Click on
view SR
And see the
remaining asssay 
If it is less the 100 test load new sr
HOW TO SEE UNIVERSAL WASH REAGENT(UWR) AND WASTE
Go to reagent manu
Click on
view UWR/Waste 
If UWR is less then 200ml then click on change both.
Put new UWR AND EMPTY WASTE
HOW TO PUT SAMPLE
Ready a sample tray

IF serum more then 1 ml then put barcoded vacutte

-
If single test is there sample volume should be more then 300 microliter
For more then 1 test sample volume should be 500 microleter in vitrous cup
PUT sample tray

And start the
green button 
HOW TO SEE THE REMAINING TIME
Click on result menu
Click on update and see all sample has been taken or not
Yellow color in front of sample id indicate sample processing on
HOW TO SEE RESULT
Go to result menu

click on update

Click on
plus sign(+) in front of sample id and see all result has came or not

In case of NA (no result) check the error message
HOW TO CHECK ERROR MESSAGE
Go to condition menu

see the error message and select it and click on
view description 
In that cause of error and it is solution is given