Do not forget to do microwell incubator reset after claning of microwell incubator , For microwell incubator reset go to Diagnostics-MEDS-INCUBATOR AND TRANSPORTS and click on Reset uW Inc
To prevent two test waste go to option and configuration menu then configure system then sample/result option keep all reprocessing menu off* but for beta-HCG all reprocessing menu on and reflex dilution on was done
For dilution of Beta-hCG for value greater than 15000 IU/L ,GO TO OPTION –CONFIGURE ASSAY–SELECT BETA-HCG –REVIEW/EDIT CONFIGURATION–REFLEX DILUTION ON AND DILUTION FACTOR 100 WAS DONE–CONFIGURE REFLEX–WRITE UPPER RANGE 15000 IN REFLEX TO SAME ASSAY
Whenever new examination will be started see host code of that examination in host code pdf and enter correctly in php adminer
To prevent failed calibration , before doing any calibration load reagent in instrument before 6 hr of calibration to settle out bubble