Check Communication
How To start Erba XL 1000
Turn on power supply(in electric board).
-
Turn on White switch(primary) on back left lower side of instrument.
-
Turn on red switch(secondary) on back left side instrument
-
Turn on computer.
Start MultiXL Software.
Login by UserID: Guest and Password: Guest
-
Prepare Sample rack
Use rack sequentially from No.1 to 20.
Remove cap of vacutte before keeping vacuette in rack.
In individuale Rack, place samples from position 1 to 5. (Position Nos are embossed on Rack TOP)
Do Not rotate the vacuette after putting in Rack.
If vacuette wrongly placed in rack, remove it and again place it correctly. (Do Not rotate in side rack to prevent damage of barcode sticker)
If sample are not Sufficient,then serum was separated in eppindorf cup(1.5mL) and cup is put on white 5 mL Tube. Then Tube with cup put in Rack.
- Programming sample for Analysis
When Decision to run a sample is taken,
Scenario -1 (No entry)
restart PC then restart machine
Connect to LIS interfacing software
tickmark sample barcode in status monitor,tickmark rack module in status monitor
Start Run
Track will move, Rack will move on infeed belt to analysis belt
Barcode will be read to get sample_id
To find test order
Q will be sent to LIS, O will be sent by LIS to machine
To find sample position, rack barcode and position of Rack will be scanned
Scenario -2 (Sample_ID entry)
Barcoded sample in 5 ml tube
Keep on rack
do/check following
Connect to LIS interfacing software
tickmark sample barcode scan in status monitor
Enter sample id in patient entry
click on Download Samples
Start Run
Scenario -3 (Sample_ID + Position entry)
sample in 5 ml tube/2 ml cup, no barcode
Keep on rack
do/check following
enter sample id
enter rack number and position to define sample position
click on Download Samples
Start Run
Scenario -4 (Sample_ID + Position + Test entry)
How To load sample rack on conveyor belt of Erba XL 1000
First prepare racks
load the rack on conveyor belt
Rack barcode must be on outer side
Rack front groov should be fit in outer rim of conveyor belt.
rack must be kept between two vertical prominent rim
NOTE: IF Conveyor Belt stop moving in between running batch, check for Rack module Tick in system parameter in setting.
Warning : Check for any metal things attached to lower side of rack magnet when Rack loading on conveyor Belt.