Sample Processing in Erba XL-1000

  1. Check Communication
    • For power supply
      • check power cable (grey colour) properly fit or not in machine.
    • For COMPORT
      • Check comport cable (blue) properly fit or not in machine.
    • for LIMS (for barcoding information)
      • Check usb cable (black) properly fit or not in machine.
    • For LIS
      • check LAN cable in CPU properly fit or not.
    • Three cable at CPU
    • Three cables at Instrument

  1. How To start Erba XL 1000
  1. Prepare Sample rack
    • Use rack sequentially from No.1 to 20.
    • Remove cap of vacutte before keeping vacuette in rack.
    • In individuale Rack, place samples from position 1 to 5. (Position Nos are embossed on Rack TOP)

- Programming sample for Analysis When Decision to run a sample is taken,

  1. Scenario -1 (No entry)
    • Barcoded sample in 5 ml tube
      • Keep on rack
        • do/check following
          • connect usb from pc-→machine, restart PC
          • select USB3 as LIS Barcode port in settings→system parameters

  1. Scenario -2 (Sample_ID entry)
  1. Scenario -3 (Sample_ID + Position entry)
  1. Scenario -4 (Sample_ID + Position + Test entry)

How To load sample rack on conveyor belt of Erba XL 1000

  1. First prepare racks
  2. load the rack on conveyor belt
    1. Rack barcode must be on outer side

  1. Rack front groov should be fit in outer rim of conveyor belt.

  1. rack must be kept between two vertical prominent rim

NOTE: IF Conveyor Belt stop moving in between running batch, check for Rack module Tick in system parameter in setting.

Warning : Check for any metal things attached to lower side of rack magnet when Rack loading on conveyor Belt.