Send Circular in all Department,Medical supritendent signed by Head of Department regarding Breakdown of Auto Analyser.Attach circular to laboratory request book.
Inform all first year resident of all department orally.
Do Entry of only Creatinine ,Glucose and ALT. If ALT comes High then do Bilirubin.
Do all Bilirubin of NICU and Do Electrolyte of only MICU,NICU,SICU,Hemodialysis unit and Critical patients.
Troubleshooting
If error massege show autospan calibration failed, look for the gain of filters, If it shows same gain of all filters (e.g.,16), it indicates lamp failure(lamp not lightening) and changing the lamp is required. How to change lamp