Clean the Bio-hazardous waste can.
Clean the Cleaning Solution can.
Ensure enough CAL A and CAL B solution in ISE.
Replace the external tubing to the Waste, Bio Hazardous Waste,Cleaning Solution and DI Water Plant.
Clean the analyzer’s external surface.
Clean the computer, monitor, keyboard and printer external surface.
Clean the area around the analyzer, and discard any unwanted item(maintain proper room cleanliness).
Check and replace the Sample and Reagent probes if necessary.9Check and replace the stirrer paddles.
Replace the laundry probes.
Clean the STAT tray.
Clean the Reagent trays.
Check and replace the syringe dilutor assembly.
Clean the fans.
Clean the bar code readers.
Perform the auto span check and note down the gain values for all thewavelengths.
Perform a cuvette rinse.
Perform a precision check and note down the %CV for an end point andkinetic test.
Replace the lamp.
Clean the internal surface free of dust.
Perform an auto span check and note down the gain values for all thewavelengths.
Carry out site verification for temperature, line voltage, electrical ground,ventilation, external interferences, room lighting, and laboratorycleanliness practice. Refer section 3.2 Site Requirementsfor moredetails.
Make a detailed entry in the error log book, of the maintenance carriedout and site verifications. Average life of the Lamp is 1000 hours