Potassium examination procedure
Learn operation of
Erba XL-640
and
Erbachem
.
Select equipment
Use Erba XL-640 as peferred equipment.
If Erba XL-640 is malfunctioning and can not be repaired within period acceptable to clinicians, use Erbachem and follow
procedure for responding to Erba XL-640 failure
Read following documents
references
K+ Reference literature 1
environmental and safety controls
MSDS of sodium chloride
MSDS of Potassium chloride
MSDS of Lithium carbonate
Preparation of
Electrolyte reagent
Preparation of Aliquots
Write sample id on ependroff cups with permanent marker pen on the rough surface of ependroff cup.
Arrange them in ascending order in the rack.
Aspirate serum from the centrifuged vacutte with the help of 200 uL pipette.
Dispense the serum into the respective sample id ependroff cup.
Recap the ependroff cup as soon as possible after dispensing serum.
Put ependroff cups with cap in refrigerator(2-8 ') if analysis is delayed.