How to make serum pool for IQC of vitros 3600

  1. First prepare serum pool of two level normal and abnormal of 50 ml in 50 ml plastic wash bottle
  2. Kept at room temperature for at least 2 hr
  3. Mix throughly and filled it in washed plain vaccutte for centrifugation
  4. make allicort of 400 microliter parafilmed it and kept in deep fridge

Frequency of IQC of VITROS-3600

  1. 1 level on every monday
  2. take two cup from deep fridge
  3. from 1 cup -ferritine,TSH,free T3,free T4 and PCT
  4. from 2nd cup -NBNP2,hsTni,FSH,LH,B-hCG and prolactin