In the morning,before daily maintainence, this wash has to be done.
Fill fresh Hypochloride in both R1(54 position ) & R2(56 position)bottles in Reagent tray.
Keep a cup filled with Hypochloride as a sample in sample tray position No-1.
Select HYPO as test and run the sample.
Remove reagent bottle and discard Hypochloride and keep the bottles in freeze no-3 at 2-8“C.
After performing this test, DAILY MAINTENANCE has to be done.
Then Reagent filling and QCs are runned.