Glucose examination procedure

Note: When a new lot is opened, verify kit literature with the uploaded version. Upload newer version when required and make necessary changes in procedure and process|

environmental and safety controls
MSDS of Transasia Kits
MSDS of randox kits

Erba XL-640

Erbachem

Note: Following instruction is useful only if operator is familiar with Erbachem

^Parameter^Value^

Mode1-POINT
Filter505 nm
Factor1
Delay T5
Read T0
Temperature37 'C
Unitmg/dl
Sample volume(ul)10
Reagent volume(ul)1000
Aspitation Volume(ul)800, Very important
Decimal place4
Calibration, Result calculation, Quality control and sample examination in Erbachem

paper on lot to lot variationpaper_on_lot_to_lot_variation.pdf

  1. Perform sample Centrifugation
  2. Take 1 ml reagent in 1.5 ml eppendorff cup in required numbers and bring to room temprature(30 min)
  3. Take 10 microliter sample in each tube at 30 second interval. Change tip every time.
  4. incubate at room temprature for 1 hour.
  5. Select “Run test” → Glucose in Erbachem menu.
  6. When instrument asks “Aspirate water”, aspirate reagent instead. (To correct for absorbance of reagent blank)
  7. When instument ask for reagent blank, press 'No' (To prevent automatic correction of factor)
  8. take reading at 505 nm with 30 second interval.
  9. Between two sample wash cuvet with water.
  10. Take care not to exceed 30 second between reading of two samples.
  11. Enter results( Actually absorbances) in spreadsheet
  12. Evaluate factors and QC results. If factor/QC results show major deviation from past, rerun batch.
  13. If sample absorbance more than 2.0 then dilute the sample about 1:1 with DI water and rerun the sample.
  14. Calculate results=(Absorbance)X(Average Factor) in spreadsheet.
  15. instructions for determining quantitative results when a result is not within the measurement interval: Dilution of sample with DI water is done in same way as autodilution done by Erba XL-640