^Erba XL 640^

identity of the equipmentErba XL-640 HOSP-EQ-P-10:at_of_erba_xl_640.pdf
manufacturer’s name, model and serial number or other unique identificationFully Automated Chemical Analyser ERBA XL 640, Sr=640051 ,Modal no;XL-640
contact information for the supplier or the manufacturerTRANSASIA BIO-MEDICALS Ltd,MR.JITESH-9327110049,MR.Manoj - 9327110109,Motor Sanjaybhai-9824176933,Mr Hiren-9327110048,To register complain call: 1800 103 8226
date of receiving and date of entering into service01/06/2011
locationClinical Biochemistry Laboratory
condition when received (e.g. new, used or reconditioned)New
manufacturer’s instructions-User manual-002_xl640_astm_host_manual.pdf-wash solution preparation
records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratoryRecord
maintenance carried out and the schedule for preventive maintenanceRecord
equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verificationsRecord
damage to, or malfunction, modification, or repair of the equipmentRecord

Genaral Instruction / Specific instruction for use in the laboratory

  1. Two files required for Erba MultiXL software as follows
    1. During Installation, when first time software started
      1. Product login: XL640
      2. Password: XL640 4I S R
      3. Select “ P” Zone
  2. Use following Sheet for recording of maintenance
  3. Procedure to run HYPOCLORIDE DAILY WASH
  4. Test Processing Sequencetest_sequence_parameters.ods
  5. Monthly Maintanence Sheet maintenance_sheet.ods

Trouble Shooting For ERBA XL 640

  1. ERROR CODE: ASP rotation error : Solution: Open front cover of ERBA XL 640. Just below sample tray look for two sensor, 1)front side and 2)on right hand side(might be lighting or not). clean the second sensor with brush. clean the plate of sensor with brush.
  2. ERROR CODE: No flow during ISE maintenance: Solution : Clean the electrode path by flushing ISE wash solution(pepsin) with syringe.

Backup and Restore of Erba XL Software

Deleting Old data from Erba XL-640 Software

    sqlcmd -S "\SQLEXPRESS"
use MultiXL
 select * from tblSamples
go

(To minimize this step of showing the datas which are going to be delete, press “ctrl+c”

delete from tblSamples
go

(Wait for some seconds)

 
 select * from tblResults
go
delete from tblResults
  (Press "ctrl+c")
*press enter
*Write following at prompt
go

MONTHLY BACKUP

-sop_for_weekly_backup.odt

*Driver for PCI comport* ch35xdrv.zip *Desk pin setup exe file in zip* deskpins130.zip