Add 2.5 ml buffer(R1a) to one bottle of R1b. Mix by gentle swirling.
Use within 3 days.
Keep at 2-8 'C when not in use.
Run control provided in the kit as shown in sample procedure. Run control with each 2.5 ml reconstitution.
Evaluate QC results. If QC results are acceptable proceed for sample examination.
Sample procedure
Select “Run test” → CK in Erbachem menu.
When instrument asks “Aspirate water”, aspirate water.
When instrument asks “Aspirate sample”.
Take 500 microliter reagent in ependroff cup
Add 10 microliter sample to it. Mix.
Aspirate reaction mixture.
Wait till reading is available.
Enter result in worklist sheet.
-
-
-
instructions for determining quantitative results when a result is not within the measurement interval: Dilution of sample with DI water is done in same way as done by Erba XL-640 automatically.