Currently, testing of CHE is done in ERBA XL-640.
The reagent is kept in the freezer of refrigerator no.11.
It has to be melted only when sample has to be tested for CHE.
Value of QC 5 (normal range) or QC 8 (abnormal range) are very similar so only one level quality control(QC 5 Normal Range) has to be used as quality control.
Sample as well as QC ( QC 5 ) is diluted 5 times, CHE is selected, testing is done.
Entry of sample is done as it is, no suffix i.e. dil is added.
QC 5 done every time whenever test done.
if more than one sample run at one time,only one qc 5 is need to be done.
Every time before analysis the sample is diluted. For that precaution should be taken. the water to be taken should be clean DI water labelled from the “frige 12” which is QC WATER.