Table of Contents
VERIFICATION
DELTA CHECK
Following should be checked in Autoverification process
VERIFICATION
Login with own ID and Password.
Go to the View button,
Enter the ID from which verification is to be started.
Click on view detail
That report will be shown.
now this report here is unverified
Click on edit button
now click the yellow color
sync ALL
button.
All values will come inside but needs calculation for calculated parameters.
Now click on calculate button.
Calculation will be done automatically as shown below.
now click on verify button and then click on verification done.
Then click on NEXT button for verification of next sample ID and repeat the procedure.
How to do autoverification
autoverification_example.pdf
DELTA CHECK
Login to LIS with your ID and Password.
Click on Print.
Click on QReport.
Following screen will be shown to you.
Enter Sample ID for which you want to do delta check.
Click on OneByOne.
Following screen will be shown to you.
This is cumulative of all parameters of single patient.
Following should be checked in Autoverification process
While batch is running, as the result of test are given by machine, start checking graph of each parameter one by one
.
Delta Check
:Compare results with previous results of same patient shown in Green highlighted form.
Watch for any
critical reports
Check whether proper remark is added or not in case of
Interference
e.g. Hemolysis,Lipemia etc
If any result columm is blank,check for following.
Results are exported or not
Quantity of sample
Transposition of sample
Sample received or not(in case of OPD)
Instrumental error e.g skiping of sample, Reagent bottle empty,
Check “Alert” columm for proper alert given or not. for example,If any value of examination is absurd , check for “High absurd” tag.