Procedure for ALT examination

referencesalt_kit_lit.pdf}}

comparison_of_alt_with_single_reagent_and_double_reagent

xl-1000_calibration_alt-lot-b122074_2021-5-20.ods

environmental and safety controls
MSDS of Transasia Kits
MSDS of spinreact kits

Preparation and use of ALT reagent 001_alt_reagent..ods

Erba XL-640

Erbachem

Note: Following instruction is useful only if operator is familiar with Erbachem

ParameterValue
Sample volume(ul)50
Reagent volume(ul)500 (R1+R2 mixed in proportion of 24 ml+6 ml)
Filter340 nm
Aspitation Volume(ul)300
Type of reactionRate A
Decimal place0 (but 4 during calibration)
Delay60 seconds
Read120 seconds
Calibration, Result calculation, Quality control and sample examination in Erbachem
Prepare for calibration as below when lot changes or there is relevent NC. Refer Calibrator and QC
testing materialReplicates
Water blank2
Normal Range Calibrator4
Abnormal range Calibrator4
Normal range QC sera4
Abnormal Range QC sera4
  1. Run all testing material as sample (as described below at Sample Examination)
  2. Evaluate calibration and QC absorbances in spreadsheet. Compare with past results.
  3. If results are acceptable, enter the factor in the erbachem and set decimal place to zero.

Sample Examination

  1. Mix R1 and R2 as described in literature in sufficient quantity.
  2. Generally 24ml R1 +6 ml R2 is mixed by adding R2 to R1.
  3. Fill 500 ul cups with mixed reagent in required numbers.
  4. Select “Run test” → GPT in Erbachem menu.
  5. When instrument asks “Aspirate water”, aspirate water.
  6. Add 50 microliter sample to a cup, mix and aspirate.
  7. If result comes more than 5000u/L then dilute the sample to 1:1 and rerun the sample.
  8. Wait and enter reading in worklist sheet.
  9. Enter sample results manually in LIS. Store the result sheet in a file for at least 2 days for use in event of LIS failures and rechecking results for NCs.
  10. instructions for determining quantitative results when a result is not within the measurement interval: Dilution of sample with DI water is done in same way as done by Erba XL-640 automatically.