====WDI OF AUTOMATED ESR ANALYZER==== __Sample Preparation:__\\ • ESR – vacuum blood collection tube contains 3.8% Sodium Citrate & vaccum to draw 1.28ml of blood.\\ • Write down patient’s lab no. on label before sample collection\\ __Procedure:__\\ • Switch on the Analyzer unit.\\ -Working time -30 min and temperature correction to 27’C is already set.\\ -Machine will display “TEST” menu automatically. If not then manually enter main menu to move cursor to “TEST” & press “OK” to enter this menu.\\ -On the screen, test tube positioning plate is highlighted in which actual positions of test tubes are described as below:\\ *e_ : This channel has no tube. *ESR : Sedimentation tube is inserted and waiting for test. *E2 : This channel is performing sedimentation testing, digitstand for scanning testing. *… : This channel is scanning testing. *h : This channel is waiting for inserting tube. *Hct : This channel is performing hematocrit testing. *8 : Sedimentation value of this channel is 8 mm/hr. *H48: Erythrocyte hematocrit is 48% • Take a ESR – vacuum blood collection tube and fill blood sample upto mark and mix it well.\\ • Introduce tube into reading channel. ESR indicator will appear on screen at respective ESR tube position.\\ • Enter the patient ID/lab no. into register according to sample position.\\ • Read the result after 30 minutes and mention it into register according to sample position.\\ • Remove the tube and discard it.\\ • Swith off the analyser unit.\\