====WDI OF AUTOMATED HEMATOLOGY ANALYSER- MICROS-60==== START: * Switch on main power supply * Switch on UPS * Switch on the cell counter by pressing the button at the rear of the instrument. * Press ID keys on front panel to enter the sample ID. * Press enter key. * Well mixed the sample. * Place the sample beneath the sample probe. * Press the manual sample bar. SHUTDOWN: * Press the stand by key * Switches to ‘stand by mode’ * Switch off the instrument * Switch off the UPS * Switch off main power supply. DAILY MAINTENANCE -Startup and standby cycles * Press the “startup” key of the front panel at the beginning of each day. * Press the “standby” key of the front panel at the end of each day. -Automatic cleaning * When an automatic cleaning cycle begins a message will appear on the screen. * A manual cleaning done by press the “auto clean” in the main menu of the screen. -General cleaning External cleaning\\ i.Slightly wet a sponge with disinfectant product and wipe the dirty surface. *Never spill liquids on the instrument *Never use disinfectant that contains alcohol. *Never use solvent or abrasive materials. *Wipe of any trace of blood spillage as soon as possible. Internal cleaning\\ ii.Concentrated cleaning *Go in the main menu, then in service, and in concentrated cleaning. *Follow the steps that appear on the displays. iii. Sampling probe *Prepare a solution of sodium hypochlorite to 100ml/l. *Fill a 5ml tube with this solution. *Run 5 analyses on bleach. iv. Backflush *Go in the main menu, then in service, then in backflush.