===== SOP for INCUBATOR===== 1.** PURPOSE:** This instrument is used for incubation of sample in Reticulocyte count test 2.** LOCATION:** Hematology Laboratory 3.** OPERATION:** • The incubator is switched on and allowed to warm up to the required temperature. • The incubator is started to use after the temperature has reached 370C. • The Samples are kept inside after the temperature has reached 370C. 4.** MAINTENANCE SCHEDULE: ** • The incubator is cleaned weekly and after spillage of infectious material. The interior of the incubator is cleaned using sterile cotton or gauze soaked in 70% Alcohol. After cleaning, the interior is allowed to dry completely. • The temperature of incubator is recorded daily in the morning and in the evening in the temperature monitoring sheet which is pasted on the incubator. The incubator should maintain a constant temperature of 350C ± 20C. • Daily temperature ,monitoring is carried out , the records of which are filed in in c\records \file\22 \Inernal maintainance records 5.** VALIDATION OF INSTRUMENTS:** To be done on installation of new Incubator & after Repair or Service. Use a calibrated thermometer to ensure accuracy of temperature checks. 6.** BACK UP PROCEDURE FOR EQUIPMENT FAILURE:** In case of failure, incubation is done in other incubators of Bacteriology laboratory. 7.** RESPONSIBILITY:** Technical staff and doctors 8.** SAFETY PRECAUTIONS:** Any operating problem should be recorded and to be brought in notice of the laboratory In-charge. Service persons are informed regarding the problem as soon as possible to solve the problem. 9.** REVIEW, RECORDS,REMARKS & RECOMMENDATION: ** • In Incubator log book No- c\records \file\22 \Internal maintenance records • Calibration details in c\records\file\21\calibration records.