====WDI of Centrifuge==== Put the power cable to socket. Turn on Switch. Open the lid of centrifuge. Load the rotor with its components of tubes/ Vacutainer. Close the lid of Centrifuge. Move speed control knob clockwise from “0’ position. To stop move speed control knob to “0’ position. Centrifuge de-accelerates. Open the centrifuge lid. Remove tubes/vacutainer **MAINTENANCE SCHEDULE** *If there is any spillage, wipe and disinfect immediately with freshly prepared (within 8 hrs.) 1 % Sodium hypochlorite. Let the hypochlorite in the bucket for 1 hr. After 1 hr, invert buckets to drain dry. Clean and Wipe. *In case of breakage of any test tube samples, immediately switch off the instrument ,open it , take it out all braked particles with gloved hands and then wipe and disinfect immediately with freshly prepared (within 8 hrs.) 1 % Sodium hypochlorite. Let the hypochlorite in the bucket for 20 minutes. After 20 minutes, invert buckets to drain dry. *Clean and Wipe again.