====WDI of Urine Quality Control Examination==== *This is done by running the BIO RAD quality control samples (level 1& level 2) daily once on auto analyzer. *Put drops of control sample level 1 on each reagent area of reagent strip with pipette. *Blot by touching the edge of strip to Tissue paper to remove excess control sample. *Insert the Reagent Strip in the Space provided for the Strip with reagent side facing upwards. *Wait for Results to appear on Screen and to have results in printed format. *Repeat procedure for control sample level 2. *The materials to discard is disposed as per the SOP of BioMedical Waste disposal *Results of control samples are compared with the reference range given with the controls *The results of daily control run are plotted in table and are filed in CP:C/records/file/9/internal Quality control records-A with the printed copies of analyzer.