=====Centrifuge===== - Insert the power cord & switch on the power supply. - RPM & time display will blink last run RPM & time. - Time setting : Push SET time key, the timer display starts incrementing till 99 & after that it starts from 1 again. Range of time – 1-99 min. Set your desirable time.
  1. RPM setting:

Push set RPM key for 3 seconds. Dot point of last RPM digit starts blinking. Now turn the RPM knob to set the desired RPM. Turning the knob clockwise - ↑se RPM. Anticlock wise - ↓ se RPM.
  1. After setting time & RPM, load test tube. Put test tubes by strictly maintaining their balance centrifuge loads must be balanced by placing the tubes of equal size, shape & containing equal volumes opposite to each other. Do not overload.

  2. Push start key,

  3. When centrifugation complete. Allow it to stop completely. Wait for 3 minutes & then open the cover of centrifuge & remove sample tubes.

  4. If during centrifuge – user wants to terminate centrifugation – Push stop key.

Routine maintenance * Interior of the bowl should be inspected regularly for contamination & SIB be cleaned regularly. * Open the lid & clean centrifuge bowl. From inner side with clean mop using neutral agent. * Remove any impurities or foreign objects that are left inside the chamber. * Remove profile rubber. Cushions from centrifuge loads & powder them with talcum.