======Procedure for Urea examination====== *Learn operation of [[Erba XL-640]] and [[Erbachem]]. *Select equipment *Use Erba XL-640 as preferred equipment. *If Erba XL-640 is malfunctioning, stop the examination and follow [[procedure for responding to Erba XL-640 failure]] *Read following documents *urea erba kit literaure{{ :nchsls:c:biochemistry:document:urea_erba_kit_literature.pdf |}} |**references **|{{Urea_kit_literature_transasia_page1.pdf|Urea Kit literature of Transasia page 1}}| |Note- 26th cycle OD of both level (qc 5, qc8) has to enter in URBL in LJ chart manually three times a day with each IQC run. 26th cycle OD should be more than 1.2, if it goes below 1.2, it is the sign of urea reagent deterioration.| | |{{Urea_kit_literature_transasia_page2.pdf|Urea Kit literature of Transasia page 2}}| | |{{urea_kit_litrature_spinreact.pdf|urea kit litrature spinreact.pdf}}| | |{{:nchsls:c:biochemistry:document:urea_kit_literature_meril.pdf|urea kit litrature meril}} | |**Note:** When a new lot is opened, verify kit literature with the uploaded version.| | |[[lis_upload_dokuwiki_media|Upload]] newer version when required and make necessary changes in procedure and process| |**environmental and safety controls**| |{{msds_transasia_kits.pdf|MSDS of Transasia Kits}}| |{{:nchsls:c:biochemistry:document:msds_spinreact_kit.pdf|MSDS of spinreact kits}}| *Document and revise [[common laboratory informations related to the examinations]]. *Depending on the selected equipment follow appropriate procedure described below *NOTE* fOR EACH TIME, ======Erba XL-640====== |[[Centrifugation]]| |[[Preparing aliquot]]| |[[nchsls:c:biochemistry:document:how_to_perform_examination_in_erba_xl-640|Procedure for examination in Erba XL-640]]|