=====Sample Processing in Erba XL-1000===== - **Check Communication** * For power supply * check power cable (grey colour) properly fit or not in machine. * For COMPORT * Check comport cable (blue) properly fit or not in machine. * for LIMS (for barcoding information) * Check usb cable (black) properly fit or not in machine. * For LIS * check LAN cable in CPU properly fit or not. * Three cable at CPU *{{:nchsls:c:biochemistry:document:wire.jpg|}} *Three cables at Instrument {{ :nchsls:c:biochemistry:document:3.jpg?600 |}} - **How To start Erba XL 1000** * Turn on power supply(in electric board). * {{:nchsls:c:biochemistry:document:main_switch.jpg?400|}} * Turn on White switch(primary) on back left lower side of instrument. * {{:nchsls:c:biochemistry:document:white_switch.jpg?400|}} * Turn on red switch(secondary) on back left side instrument * {{:nchsls:c:biochemistry:document:red_switch.jpg?400|}} * Turn on computer. * Start MultiXL Software. * Login by UserID: Guest and Password: Guest * {{:nchsls:c:biochemistry:document:login_id.jpg|}} -** Prepare Sample rack** * Use rack sequentially from No.1 to 20. * Remove cap of vacutte before keeping vacuette in rack. * In individuale Rack, place samples from position 1 to 5. (Position Nos are embossed on Rack TOP) {{:nchsls:c:biochemistry:document:img_20200604_131715.jpg?600|}} * Place the Vacuette with barcode on front side in Rack. {{:nchsls:c:biochemistry:document:front_side_of_rack.jpg?400|}} {{:nchsls:c:biochemistry:document:back_side_of_rack.jpg?400|}} * Do Not rotate the vacuette after putting in Rack. * If vacuette wrongly placed in rack, remove it and again place it correctly. (Do Not rotate in side rack to prevent damage of barcode sticker) * __If sample are not Sufficient,then serum was separated in eppindorf cup(1.5mL) and cup is put on white 5 mL Tube. Then Tube with cup put in Rack.__ - **Programming sample for Analysis** **When Decision to run a sample is taken**, *Three parameters are required * Sample id * Sample Position * Sample test Order -__Scenario -1 (No entry)__ *Barcoded sample in 5 ml tube *Keep on rack *do/check following *connect usb from pc-→machine, restart PC *select USB3 as LIS Barcode port in settings→system parameters {{:nchsls:c:biochemistry:document:com3_in_softwr_.jpg?600|}} {{:nchsls:c:biochemistry:document:comb-3.jpg?600|}} {{ :nchsls:c:biochemistry:document:all2.jpg?600 |}} *restart PC then restart machine *Connect to LIS interfacing software *tickmark sample barcode in status monitor,tickmark rack module in status monitor {{ :nchsls:c:biochemistry:document:eallc2.jpg?600 |}} *Start Run *Track will move, Rack will move on infeed belt to analysis belt *Barcode will be read to get sample_id *To find test order *Q will be sent to LIS, O will be sent by LIS to machine *To find sample position, rack barcode and position of Rack will be scanned -__Scenario -2 (Sample_ID entry)__ *Barcoded sample in 5 ml tube *Keep on rack *do/check following *Connect to LIS interfacing software *tickmark sample barcode scan in status monitor *Enter sample id in patient entry *click on Download Samples *Q will be sent to LIS, O will be sent by LIS to machine to get test order *Start Run *Track will move, Rack will move on infeed belt to analysis belt *Barcode will be read *To find sample position, barcode and position of Rack will be scanned -__Scenario -3 (Sample_ID + Position entry)__ *sample in 5 ml tube/2 ml cup, no barcode *Keep on rack *do/check following *Connect to LIS interfacing software *enter sample id *enter rack number and position to define sample position *click on Download Samples *Q will be sent to LIS, O will be sent by LIS to machine to get test order *Start Run -__Scenario -4 (Sample_ID + Position + Test entry)__ *sample in 5 ml tube/2 ml cup, no barcode *Keep on rack *enter sample id *enter rack number and position to define sample position *enter tests ordered *Start Run **How To load sample rack on conveyor belt of Erba XL 1000** - First prepare racks - load the rack on conveyor belt - Rack barcode must be on outer side {{:nchsls:c:biochemistry:document:rack_last.jpg?400|}} - Rack front groov should be fit in outer rim of conveyor belt. {{:nchsls:c:biochemistry:document:front_groove1.jpg?600 |}} {{:nchsls:c:biochemistry:document:rack_front_groove.jpg?400|}} - rack must be kept between two vertical prominent rim {{:nchsls:c:biochemistry:document:rack_first.jpg?400|}} {{:nchsls:c:biochemistry:document:rack_side.jpg?400|}} * If Instrument is in running status, - load all the samples in rack, prepare racks - Pause the instrument from status monitor - load the sample racks on conveyor belt - resume run **NOTE:** IF Conveyor Belt stop moving in between running batch, check for Rack module Tick in system parameter in setting. **Warning** : Check for any metal things attached to lower side of rack magnet when Rack loading on conveyor Belt. {{ :nchsls:c:biochemistry:document:magnet.jpg?400 |}}