1. Starting Power -Switch on main(wall) supply -Start UPS supply by pressing ON-button for few seconds -Switch on Erba XL-640 back-power-supply (Wait for 2 min) -Do not Switch on Erba XL-640 side-power-supply -(side-power-supply=OFF > Front light=RED) -(side-power-supply=ON > Front light=GREEN) -Switch on CPU and monitor power-supply -If UPS gives audio alarm > call resident 2. Starting software -Start software by double clicking “MultiXL” icon on Desktop -Login window will appear. -Enter -login id = tech -Password = 12345678 then click enter. 3. Reagents placement -Do not Switch on Erba XL-640 side power-supply during reagent refilling. -See the reagent tray to see which reagents need to be refilled. -Refill reagents twice a day, in the morning and evening using [[loading reagents in Erba XL-640]] -Place ISE cleaning solution(1 ml) at ISE-2 position of sample wheel 4. Program first run of morning -Do not Switch on Erba XL-640 side-power-supply during programming a run. -In patient entry=> write sample id of qc 5 as 5YYMMDDTT eg. 514062109 then check the group and position.similarly give entry of qc 8 .Then click on QC morning and then save . For MPR QC 5 , give entry of MPR and CGLC .and then save. {{patient_entry 1.jpg|Patient entry}} 6. Placement of QC(0.5 ml) -Do not Switch on Erba XL-640 side-power-supply during positioning QC samples -Bring at room temperature and mix -Keep QC-5 at position-1 -Keep MPR-5 at position-2 -Keep QC-8 at position-3 -Keep MPR-8 at position-4 -Make sure that you entered the samples in that sequence in patient entry. 7. Placement of Samples(0.4 ml) -Do not Switch on Erba XL-640 side-power-supply during positioning patient samples -Open [Patient Entry]--->[Worklist] and place samples at programmed position 8. Starting run of morning -Switch on Erba XL-640 side-power-supply NOW. -side-power-supply=ON > Front light=GREEN -Daily Morning maintenence -Do maintainance as follows:{{maintainance.jpg|Maintainance}} - Click Prime wash,it takes around 10 minutes. On completion, message is displayed that Prime wash successfully completed. - Click on R1/R2 Probe wash.Message will be displayed "put washing solution at position 55 in reagent tray and press OK to continue" click OK if it is present , if not then put the solution at that position and then click OK.It will take around 3 minutes.On completion, message is displayed that R1/R2 Probe wash successfully completed. - Click on Sample Probe wash.Message will be displayed "put cleaning solution at ISE2 position in sample tray and press OK to continue" click OK if it is present , if not then put the solution at that position and then click OK..It will take around 3 minutes.On completion, message is displayed that sample Probe wash successfully completed. -Click on ISE ON=> message will be displayed "Please wait"=> Click on Clean => then click start.On completion a green box will be displayed. then again click start.A new green box will appear on completion. -Click Purge A => Click start => On completion a green box will be displayed. then again click start.A new green box will appear on completion.=> repeat 4 times. -Press [Status Monitor]-->START button to start a run -When Run is going on, START button is GRAY. -Do not Open Front cover/place sample/place reagent till START button is gray. -When run is complete, START button is GREEN again. -Switch OFF Erba XL-640 side-power-supply -Make sure that Front light=RED -Open front cover, remove samples, close front cover. 9. Using Erba XL-640 after first run of morning: -Program samples:(Only when front light is RED) -Place samples:(Only when front light is RED) -Switch on Erba XL-640 side-power-supply (front light is GREEN) -Press [Status Monitor]-->START button to start a run -When Run is going on, START button is GRAY. -Do not Open Front cover/place sample/place reagent till START button is GREEN. -When run is complete, START button is GREEN again. -Switch OFF Erba XL-640 side-power-supply -Make sure that Front light=RED -Open front cover, remove samples, close front cover. 10. Fluid levels check: ^Container^Action^Indication^ |DI Water Container|Fill upto upper mark|When level reach lower mark| |Waste|Empty upto lower mark|When level reach upper mark| |Cleaning solution|Replace cleaning solution Container|When level reach lower mark| |Biohazardous waste|Empty upto lower mark and add 2 liter 4-6% HOCl|When level reach upper mark| Make cleaning solution by adding 100 ml of Erba XL-Wash/Extran in 10 liter water 11. When -Alarm bell rings and -Message “Check Fluid Level” is shown in error message window -Then Check level of following containers and take action: ^Step^Container^Action^ |1|DI Water Container|DIW fill upto mark|If Alarm do not stop > go to step-2| |2|Waste|Empty upto mark|If Alarm do not stop > go to step-3| |3|Cleaning solution|DIW fill upto mark + 100ml stock cleaning solution|If Alarm do not stop > go to step-4| |4|Biohazardous waste|Empty upto mark,then add [2 L HOCl]|If Alarm do not stop > call resident|