====Daily Maintenance==== **//Start of the day procedure//** - Empty Bio-hazardous Waste can. - Fill the Cleaning Solution can. - Clean the analyzer external surface. - Replace the printer paper if, necessary - Clean the Computer, Monitor, Keyboard And Printer External Surface. - Perform Prime, Cuvette Rinse and Probe Wash operations and Check CellBlanks (All these operation will be performed automatically when theinstrument is switched on first time). - Clean all the probes from outside with Alcohol. - Verify Reaction Tray & RGT1 & RGT2 trays Temperatures. - Replenish or Replace Reagents If Necessary. - Perform ISE Cleaning Once and ISE Purge Commands 3-4 times. - Ensure enough CAL A and CAL B in ISE. - Remove the Reagent Trays RGT1 & RGT2 and clean the water condensedon the tray and Reagent Bottles using a dry tissue //**End of the day Procedure**// - Remove and discard all Sample, Standard, and Controls cups or tubesfrom the STAT tray. - Carry out Auto Wash if latex based chemistries are used during the day. - Perform the water save. - Remove any condensation present on the walls of the Reagent Trays RGT1 & RGT 2 or near the Reagent Bottle necks using a dry tissue. - Empty the bio-hazardous waste can. - Clean analyzer external surface, to remove residues of serum, reagentsetc. ====Weekly Maintenance==== - Clean and fill the Bio-hazardous waste can. - Clean the Cleaning Solution can. - Ensure enough CAL A and CAL B solution in ISE. - Clean the stirrer paddles and laundry probes with alcohol. - Check the syringe assembly and surrounding tubes. - Clean the computer, trolley, monitor, keyboard and printer externalsurface. - Clean the area around the analyzer, and discard any unwanted item.(Maintain proper room cleanliness). - Clean the Sample and Reagent probes. - Clean the STAT tray. - Clean the Reagent trays RGT1 and RGT2.11Perform an auto span check and note down the gain values for all thewavelengths. - Perform a cuvette rinse and check cuvette blanks. - Clean the barcode reader window, in the piece of RGT base unit. - Archive MultiXL database - Backup MultiXL database – Main Database, Archive database ===Annual Maintenance=== - Clean the Bio-hazardous waste can. - Clean the Cleaning Solution can. - Ensure enough CAL A and CAL B solution in ISE. - Replace the external tubing to the Waste, Bio Hazardous Waste,Cleaning Solution and DI Water Plant. - Clean the analyzer’s external surface. - Clean the computer, monitor, keyboard and printer external surface. - Clean the area around the analyzer, and discard any unwanted item(maintain proper room cleanliness). - Check and replace the Sample and Reagent probes if necessary.9Check and replace the stirrer paddles. - Replace the laundry probes. - Clean the STAT tray. - Clean the Reagent trays. - Check and replace the syringe dilutor assembly. - Clean the fans. - Clean the bar code readers. - Perform the auto span check and note down the gain values for all thewavelengths. - Perform a cuvette rinse. - Perform a precision check and note down the %CV for an end point andkinetic test. - Replace the lamp. - Clean the internal surface free of dust. - Perform an auto span check and note down the gain values for all thewavelengths. - Carry out site verification for temperature, line voltage, electrical ground,ventilation, external interferences, room lighting, and laboratorycleanliness practice. Refer section 3.2 Site Requirementsfor moredetails. - Make a detailed entry in the error log book, of the maintenance carriedout and site verifications. Average life of the Lamp is 1000 hours ====Quarterly Maintenance==== - Clean the bio-hazardous waste can. - Clean the Cleaning Solution can. - Ensure enough CAL A and CAL B solution in ISE. - Clean the computer, monitor, keyboard and printer external surface. - Clean the area around the analyzer, and discard any unwanted item(maintain proper room cleanliness). - Clean the Sample and Reagent probes. - Clean the Stirrer paddles.8Clean the laundry probes. - Clean the STAT tray - Clean the Reagent trays. - Clean the fans. - Clean the bar code reader. - Clean the Waste hub. - Perform the auto span check and note down the gain values for all thewavelengths. - Perform a cuvette rinse. - Replace the lamp.17Clean the internal surface free of dust. - Perform an auto span check and note down the gain values for all thewavelengths. - Carry out site verification for temperature, line voltage, electrical ground,ventilation, external interferences, room lighting, and laboratorycleanliness practice. Refer section Pre-installation requirementsfor moredetails. - Make a detailed entry of the maintenance carried out and site verificationsperformed in the error log book. - Cleanup hard disk for Temporary files like old communication log files by archiving. ====**Preventative Maintenance**==== As per manufacturer's user manual.