=====Special instruction for hs Tn-I calibrator==== -3 level of calibrator level-1 , level-2, and level-3 -It comes in frozen state and ready to use form -Each level contain 2 ml of calibrator -From main bottle 4 cup was made of 0.5 ml from each level -This cup was capped and parafilm done -Barcode label was sticked on lower part which has mentioned Name lot no and level - kept in urine micro-container and labelled as **fresh cup** -For next time calibration use this cup -After calibration retained calibrator was kept in original bottle in which it actually came and it was labelled as **used cup** -This retained calibrator will be used as a quality control sample in future -All this kept in box -This box kept in deep friedge no-16 -** After each calibration kept retained calibrator in its original bottle** -** For calibration do not use sample from its original bottle ,use labelled parafilmed cup** -[[tropi,PCT,proBNP mean and sd data]] -[[Kit literature of hS TROP I test]] ======VITROS 3600====== |[[Centrifugation, Preparing aliquot, how_to_perform_examination_in_VITROS 3600]]| |[[INTERNAL QUALITY CONTROL]]| {{ :nchsls:c:biochemistry:document:lot_verification_of_vitros-3600.ods |}} |[[EQA For VITROS_3600 Chemianalyzer Tests]]|