**Note:** Following instruction is useful only if operator is familiar with [[Erba XL-640]] ^Program [[Erba XL-640]] for the examination.^ |{{:nchsls:c:biochemistry:document:test_parameters.ods|sorted test parameters}}| |{{:nchsls:c:biochemistry:document:multixllog.zip|exported test parameters}}| [[nchsls:c:biochemistry:document:calibration_and_quality_control_frequency|Current calibration frequency policy]] ^[[Important notes]]^ ^[[ Abnormal Alert and Reference Range ]]^ ^[[Management of reagent aliquot]]^ ^[[loading reagents in Erba XL-640]]^ ^[[nchsls:c:biochemistry:document:how_to_calibrate_on_erba_xl-640|Calibration]]^ ^[[nchsls:c:biochemistry:document:internal_quality_control|Internal Quality Control]]^ ^[[how_to_do_external_quality_control_on_erba_xl-640|EQA]]^ -[[Run samples in Erba XL-640]] -[[lis_export_result_from_erba_xl_640|Export results from Erba XL-640]] -[[lis_import_result_from_erba_xl_640|Import results in LIS]] **Preparation of Aliquots** - Write sample id on ependroff cups with permanent marker pen on the rough surface of ependroff cup. - Arrange them in ascending order in the rack. - Aspirate serum from the centrifuged vacutte with the help of 200 uL pipette. - Dispense the serum into the respective sample id ependroff cup. - Recap the ependroff cup as soon as possible after dispensing serum. - Put ependroff cups with cap in refrigerator(2-8 ') if analysis is delayed.