|**Do in house of calibration of centrifuge every 6 monthly**| ====How To Operate Remi Centrifuge PR 23==== |{{current_user manual_centrifuge_PR-23.zip|User manual of PR-23}}| |It was seen that During centrifugation vacuttee cap touches rotor of centrifuge while it is in horizontal position so it prevent the bucket to come in vertical position.to prevent this following should be done.during cotton filling equal amount of cotton should be fill in both the rotor(opposite one)in same position.see below picture:| |{{:nchsls:c:biochemistry:document:upperview_of_centrifuge.jpg?300|}}| STEPS: -Load balanced samples and close the lid -After power on check following on Display[PRG. No=0] [RPM=3000] [Rotor No=19] [Brake=M] [Min=10] -Press START button -Wait till alarm (dip noice) for 5 times and again RPM = 3000 -Open the lid and unload samples Precautions: -Do not change the parameters of programs -Press the button of lid lock of cover at the time of closing cover to prevent malfunction of lid lock -Match bucket number with number on rotor,both numbers facing each other. -Do not handle without gloves. -Stay away during centrifugation. -Switch off the centrifuge when not in use. **How To Operate Remi Centrifuge R-8C-BL ** |{{current_user manual_centrifuge_PR-23.zip|User manual of R-8C-BL}}| STEPS: -Make sure that POWER is on and MOTOR is OFF. -Load balanced samples and close the lid gently. -Check RPM speed 3000 on display by pressing RPM button. -Set time 10 min . -Turn on motor button. -Centrifuge will automatically stop at end of 10 min. -Turn OFF MOTOR. -Open the lid when display shows 0 RPM speed. Precautions: -Do not handle without gloves. -Stay away during active centrifugation. -Check rotor screws every Monday in the morning. -If lid is not closed gently, LID lock may malfuction. -Always switch off the centrifuge when not in use. -Switch off the centrifuge when not in use. **How To Operate Remi Centrifuge R-8C-DX ** STEPS: -Load balanced samples and close the lid gently with pressing lid cover button. -set time 10 min in display. -set RPM speed 3000 on display. -Do not open lid cover until display show 0 speed. Precaution: -Do not handle without gloves. -Stay away during active centrifugation. -Check rotor screws every Monday in the morning. -Switch off the centrifuge when not in use.