======HDL cholesterol examination procedure====== *Learn operation of [[Erba XL-640]] and [[Erbachem]]. *Select equipment *Use Erba XL-640 as peferred equipment. *If Erba XL-640 is malfunctioning and can not be repaired within period acceptable to clinicians, use Erbachem and follow [[procedure for responding to Erba XL-640 failure]] *Read following documents ^References|{{DS-MG2+ HDL.pdf|HDL reference literature}}| ^**environmental and safety controls**|{{msds_dextran sulphate.pdf|MSDS of dextran sulphate}}| | |{{msds_Mgcl2.pdf|MSDS of Mgcl2}}| | |{{msds of Hcl.pdf|MSDS of Hcl}}| {{ :nchsls:c:biochemistry:document:hdl_cholesterol_precipitator_reagent_new.ods |}} ======Erba XL-640====== *{{:nchsls:c:biochemistry:document:wdi_for_hdl-cholesterol.odt|WDI for HDL-cholesterol}} *Multiply Cholesterol factor by 1.1|{{:nchsls:c:biochemistry:document:hdl_dilution_factor_calculation.ods|HDL dilution factor calculation}}| to arrive at HCL-C factor. *Due to 200 ul sample + 20 ul ppt reagent dilution *If any sample has more than 90mg/dl perform test with half dilution *It has been observed that abnormal range HDL Qc sample have more than 90mg/dl result. so,dilute it before performing abnormal QC. *Recently,calibration factor is used.Factor is derived by using Randox calibrators. |[[Centrifugation]]| |[[Sample preparation for HDL]]| |[[nchsls:c:biochemistry:document:how_to_perform_examination_in_erba_xl-640|Procedure for examination in Erba XL-640]]|