[[Preparation before doing procedure for ELISA]] =====Preparation before starting free T4 ELISA===== -Dispense 100uL samples in wells of dummy plate according to worklist using 100 uL pipette. -Take out the microtiter plate. Don't touch the lower surface of the plate. -Dispense 25 uL sample from the dummy plate into the microtiter plate using multichannel pipette.Note the time of starting dispensing of sample. -Take enzyme conjugate in trough. -Dispense 100 uL enzyme conjugate in microtiter plate using multichannel pipette.Note time after last column is dispensed with conjugate. -Mix the plate in incubator- shaker for 10 sec -Incubate the plate at 24'C in the incubator for 1 hr from midpoint of start time of sample dispensing and end of conjugate dispensing. -Wash buffer:dilute 1 volume of wash buffer(25X)with 24 volumes of DI water. -To make 500 ml,Take 20 ml of wash buffer and make upto 500 ml with DI water. -After 1 hour , take the plate from the incubator. And wash the plate in the [[ELISA washer]]. -After washing, Pat dry the plate on tissue paper. -Take TMB(Tetramethyl benzedene) substrate in the trough. -Dispense 100 uL TMB substrate in the microtiter plates with the multichannel pipette. -Mix in incubator cum shaker for 10sec at room temperature. -Incubate at 24'C in the incubator for 20 minutes. -Take stop solution in the trough. -Dispense 150 uL stop solution in the microtiter plate. -Mix in incubator cum shaker for 10 sec. -Incubate for 5 minutes in the incubator at 24'C. -Take the plate from incubator and put it in the ELISA reader. -Connect ELISA reader with the computer and set Filter i.e at 450 nm . then read plate for 3 times. -Take average of the 3 OD . and sort the data and calculate results from the graph derived from Standard. -Whenever result of any sample comes negetive,it's OD will be more than OD of "0" standard OD.So,result will come negetive.So,release the result as "<0"pmol/L. -Whenever OD of any sample comes less than OD of highest Free T4 calibrator,release the result as ">value of highest calibrator".e.g.If highest calibrator value is 95pmol/L,than release the report as ">95"pmol/L. **General procedure for [[ELISA READER]]** **[[Note on using electronic pipette for ELISA]]**