====== ** Instruction for Free T3 calibrator ** ====== - 3 level of calibrator ,frozen powered form - keep at room temperature for 10-15 min - reconstitute with 1 ml of QC water - Make 5 cup of 200 microliter of each level - 1 cup of each l;evel will be used for calibration - Remained 4 cup of each calibrator label it and barcode it and place in a urine microcontained and label those as a fresh cup and use this for next time calibration - retained sample of calibration will be kept in its original bottle ,parafilmed it and place in urine microcontained and label of used cup - Kept all these cup used as well as fresh in friedge no-16 -{{ :nchsls:c:biochemistry:document:ft3_gem1021_ww_en_i_12_1.pdf |}}[kit literature of FREE T3 ] ======VITROS 3600====== |[[Centrifugation, Preparing aliquot, how_to_perform_examination_in_VITROS 3600]]| |[[INTERNAL QUALITY CONTROL]]| |[[EQA For VITROS_3600 Chemianalyzer Tests]]| {{ :nchsls:c:biochemistry:document:lot_verification_of_vitros-3600.ods |}}