**^Erba XL 640^** |identity of the equipment|Erba XL-640 HOSP-EQ-P-10:{{:nchsls:c:biochemistry:document:at_of_erba_xl_640.pdf|at_of_erba_xl_640.pdf}}| |manufacturer’s name, model and serial number or other unique identification|Fully Automated Chemical Analyser ERBA XL 640, Sr=640051 ,Modal no;XL-640| |contact information for the supplier or the manufacturer|TRANSASIA BIO-MEDICALS Ltd,MR.JITESH-9327110049,MR.Manoj - 9327110109,Motor Sanjaybhai-9824176933,Mr Hiren-9327110048,To register complain call: 1800 103 8226| |date of receiving and date of entering into service|01/06/2011| |location|Clinical Biochemistry Laboratory| |condition when received (e.g. new, used or reconditioned)|New| |manufacturer’s instructions|-{{:nchsls:c:biochemistry:document:001_xl_640_om.pdf|User manual}}-{{:nchsls:c:biochemistry:document:002_xl640_astm_host_manual.pdf|002_xl640_astm_host_manual.pdf}}-{{:nchsls:c:biochemistry:document:003_erba_xl-640_wash_solution_.jpeg?linkonly|wash solution preparation}}| |records that confirmed the equipment’s initial acceptability for use when equipment is incorporated in the laboratory|Record| |maintenance carried out and the schedule for preventive maintenance|Record| |equipment performance records that confirm the equipment’s ongoing acceptability for use;Such records include copies of reports/certificates of all calibrations and/or verifications|Record| |damage to, or malfunction, modification, or repair of the equipment|Record| **Genaral Instruction / Specific instruction for use in the laboratory** - [[Procedure for ERBA XL-640 Software Installation]] - Two files required for Erba MultiXL software as follows - {{:nchsls:c:biochemistry:document:erba:databaseutility.zip|Database Utility}} - {{:nchsls:c:biochemistry:document:erba:setup.zip|SetUp}} - During Installation, when first time software started - Product login: XL640 - Password: XL640 4I S R - Select " P" Zone - [[Procedure for Erba XL-640 wash solution preparation]] - [[Erba XL-640 fluid managing]] - [[Procedure after Erba XL-640 probe injury]] - [[Procedure for operating Erba XL-640]] - [[Procedure of ISE cleaning solution]] - [[Procedure of changing the Lamp]] - [[Persons authorised to operate ERBA-XL 640]] - [[test parameter settings]] - Use following Sheet for recording of maintenance -Procedure to run [[HYPOCLORIDE DAILY WASH]] -Test Processing Sequence{{:nchsls:c:biochemistry:document:test_sequence_parameters.ods|}} -[[Sudden fall in Lamp Gain]] -Monthly Maintanence Sheet {{ :nchsls:c:biochemistry:document:maintenance_sheet.ods |}} -[[For pepsin wash in xl-640 breakdown ]] ===Trouble Shooting For ERBA XL 640=== - **ERROR CODE: ASP rotation error** : Solution: Open front cover of ERBA XL 640. Just below sample tray look for two sensor, 1)front side and 2)on right hand side(might be lighting or not). clean the second sensor with brush. clean the plate of sensor with brush. - **ERROR CODE: No flow during ISE maintenance**: Solution : Clean the electrode path by flushing ISE wash solution(pepsin) with syringe. ===Backup and Restore of Erba XL Software=== ===Deleting Old data from Erba XL-640 Software=== *close/shut down multixl software* *Open command prompt from Start->All Programs -> Accessories-> Command Prompt *Enter command as follows sqlcmd -S "\SQLEXPRESS" *press enter. *you will see **1>** *See screenshot * Write follwoing command at >1 prompt use MultiXL *press enter *Write folowing at prompt select * from tblSamples * Press enter *Write following at prompt go *press enter *It will show all samples in database *to stop display of long list press Control + C *Write following at prompt (To minimize this step of showing the datas which are going to be delete, press "ctrl+c" delete from tblSamples *press enter *Write following at prompt go *press enter *It will delete all samples in database *Write folowing at prompt (Wait for some seconds) select * from tblResults * Press enter *Write following at prompt go *press enter *It will show all results in database *Write folowing at prompt delete from tblResults (Press "ctrl+c") *press enter *Write following at prompt go *press enter *It will delete all results in database ===MONTHLY BACKUP=== -{{{:nchsls:c:biochemistry:document:sop_for_weekly_backup.odt|}} *for exiting , type in prompt exit twice ***Driver for PCI comport*** {{ :nchsls:c:biochemistry:document:ch35xdrv.zip |}} ***Desk pin setup exe file in zip*** {{ :nchsls:c:biochemistry:document:deskpins130.zip |}}