================== Following troubles may come ================== - Machine will show dirty cuvette errors continuously. - Machine will show Autospan failed. - Machine will stop working in between the run. - Machine will not start. ================== What to do when above listed troubles occurs ??? ================== - Start cuvette ringe and observe the flow. - Check for tube blockages. Tubes are being shown here in image{{:nchsls:c:biochemistry:document:whatsapp_image_2021-08-25_at_4.47.31_pm.jpeg?1500|}} - Check for water overflow/spills. Overflow during cuvette ringe is being shown here in video{{ :nchsls:c:biochemistry:document:whatsapp_video_2021-08-25_at_4.52.46_pmmm.mp4 |}} - Check for lamp and lamp gain. - Clean path between cuvette and lamp with histopathology comb wrapped with clean tissue paper.