==============Procedure For Sorting and Analyzing monthly Internal Quality Control and External Quality Control Results============ -Open 10.207.3.240/ -{{:nchsls:c:biochemistry:document:01.png|}} -click on LIS. Following Window will be seen. {{:nchsls:c:biochemistry:document:clickonlis.png|}} -Log in with your ID and password. Following Window will be seen. {{:nchsls:c:biochemistry:document:00002.png|}} -Now click on QC. Following Window will be seen. {{:nchsls:c:biochemistry:document:1.png|}} -Click on LJ Chart. Following Window will be seen. {{:nchsls:c:biochemistry:document:2.png|}} -Click on Date wise. Following Window will be seen {{:nchsls:c:biochemistry:document:2.png|}} - Name equipment name in box of "QC Equipment" for example XL_1000 or XL_640 or VITROS_3600. - Click on BI - following window will open. - {{:nchsls:c:biochemistry:document:3.png|}} - select test from morning QC profile seen as below. - {{:nchsls:c:biochemistry:document:4.png|}} - click on export. - one csv file will form. save it as odt file. for example {{ :nchsls:c:biochemistry:document:creatinine.ods |}} - arrange it as shown below. - {{:nchsls:c:biochemistry:document:arrange_data_form.png|}} - {{:nchsls:c:biochemistry:document:arrange_2.png|}} - - - - -Now calculate average, SD and CV of monthly data.For calculating average write in the cell "=AVERAGE( then select the data of which average is to be calculated . similarly for SD write "=STDEV(" and then follow the same as that for average. CV = SD/AVERAGE*100{{{{:nchsls:c:biochemistry:document:3333.png|}}}} - From the RIQAS file write method wise , obtained and instrument wise result. - {{:nchsls:c:biochemistry:document:4444.png|}} - And then write literature value of that lot of QC.